![ZOLL X Series Скачать руководство пользователя страница 217](http://html1.mh-extra.com/html/zoll/x-series/x-series_operators-manual_4055648217.webp)
Advisory Defibrillation Procedure
9650-001355-01 Rev. M
X Series Operator’s Guide
16-5
When a shockable rhythm is detected (ventricular fibrillation or wide-complex tachycardia
with heart rate > 150), the unit will display the
SHOCK ADVISED
message.
Press the
CHARGE
button, and press the
Select Energy
arrows up or down to select the desired energy
level if needed. Press
CHARGE
again to confirm.
Regardless of the analysis result, the user can control the defibrillator manually. For example,
the user can defibrillate the patient even if the advisory function issues a
NO SHOCK
ADVISED
message.
3 Press SHOCK button
WARNING! Warn all persons in attendance of the patient to STAND CLEAR prior to defibrillator
discharge.
Do not touch the bed, patient, or any equipment connected to the patient during
defibrillation. A severe shock can result. Do not allow exposed portions of the patient’s
body to come in contact with metal objects, such as a bed frame, as unwanted pathways
for defibrillation current may result.
Press and hold the
SHOCK
button
on the front panel until energy is delivered to the
patient.
The delivered energy level is displayed at the bottom of the screen and the shock number (1)
displays at the top of the screen and in the Defib Control panel at the bottom of the screen.
Note:
If at any time you want to cancel the defibrillation, press the
Disarm
softkey.
If the defibrillator is not discharged within 60 seconds after reaching the selected
energy level, the unit automatically disarms itself.
If additional countershocks are needed, follow steps 1 through 3 of this procedure starting on
page 16-3, to readjust the energy settings, charge the unit, and deliver shock.
Perform CPR
Begin chest compressions and rescue breathing per local protocol.
Repeat Analysis
Press the
ANALYZE
button to restart an ECG analysis and determine if additional shocks are
required.
Note:
Reanalysis of the ECG rhythm is inhibited for 3 seconds after each shock.
Continue Patient Care
Continue patient care according to medical protocols.
1
Содержание X Series
Страница 1: ...Includes Real CPR Help and See Thru CPR X Series Operator s Guide 9650 001355 01 Rev M...
Страница 64: ...Chapter 2 Product Overview 2 24 www zoll com 9650 001355 01 Rev M...
Страница 74: ...Chapter 3 Monitoring Overview 3 10 www zoll com 9650 001355 01 Rev M...
Страница 80: ...Chapter 4 Trends 4 6 www zoll com 9650 001355 01 Rev M...
Страница 90: ...Chapter 5 Alarms 5 10 www zoll com 9650 001355 01 Rev M...
Страница 110: ...Chapter 7 Monitoring Respiration Resp and Heart Rate HR 7 10 www zoll com 9650 001355 01 Rev M...
Страница 126: ...Chapter 8 Monitoring Non Invasive Blood Pressure NIBP 8 16 www zoll com 9650 001355 01 Rev M...
Страница 156: ...Chapter 10 Pulse CO Oximetry SpO2 10 18 www zoll com 9650 001355 01 Rev M...
Страница 166: ...Chapter 11 Monitoring Invasive Pressures IBP 11 10 www zoll com 9650 001355 01 Rev M...
Страница 198: ...Chapter 14 12 Lead ECG Interpretive Analysis 14 16 www zoll com 9650 001355 01 Rev M...
Страница 218: ...Chapter 16 Advisory Defibrillation 16 6 www zoll com 9650 001355 01 Rev M...
Страница 224: ...Chapter 17 Analysis CPR Protocol Defibrillation 17 6 www zoll com 9650 001355 01 Rev M...
Страница 234: ...Chapter 18 External Pacing 18 10 www zoll com 9650 001355 01 Rev M...
Страница 242: ...Chapter 19 Real CPR Help 19 8 www zoll com 9650 001355 01 Rev M...
Страница 254: ...Chapter 21 Patient Data 21 6 www zoll com 9650 001355 01 Rev M...
Страница 338: ...Specifications A 38 www zoll com 9650 001355 01 Rev M...