Chapter 18 External Pacing
18-2
www.zoll.com
9650-001355-01 Rev. M
External Pacing
X Series defibrillators contain a pacemaker for the emergency treatment of hemodynamically
compromising bradycardia, bradycardia with escape rhythms that are unresponsive to
pharmacological therapy, refractory tachycardia (supraventricular or ventricular) and
bradyasystolic cardiac arrest.
Proper demand pacing requires a reliable, high quality surface ECG signal. For best results
apply both standard ECG monitoring electrodes and hands-free pacing therapy electrodes to the
patient.
Pacer Modes
The X Series has two pacer mode settings: Demand and Fixed. The default factory mode setting
is Demand.
In Demand mode, pacing pulses are inhibited by the patient’s QRS complexes that occur during
an interval that is dependent on the setting of the rate control. If during that interval no QRS
complexes are detected, a pacing pulse is delivered to the patient. In the demand mode, the
pacer supplies the required number of pacing pulses to maintain the patient’s heart rate at
approximately the rate selected in the pacing rate window. See the procedure below, “Pacing in
Demand Mode” for more information.
In Fixed mode, pacing pulses are not dependent on the patient’s cardiac activity. Fixed pacing
should be performed only in an emergency when no alternative is available. The pacer delivers
pacing pulses at the selected pacing rate. See “Pacing in Fixed Mode” on page 18-6 for more
information.
Pacing in Demand Mode
Determine Patient Condition and Provide Care Following Local
Medical Protocols
Prepare the Patient
Remove all clothing covering the patient’s chest. Dry chest if necessary. If the patient has
excessive chest hair, clip it to ensure proper adhesion of the electrodes.
1 Turn On Unit
Press the green power switch located on the top of the unit. The green, yellow, and red lights at
the top of the unit flash on and off, and the unit displays the message SELF TEST PASSED.
Содержание X Series
Страница 1: ...Includes Real CPR Help and See Thru CPR X Series Operator s Guide 9650 001355 01 Rev M...
Страница 64: ...Chapter 2 Product Overview 2 24 www zoll com 9650 001355 01 Rev M...
Страница 74: ...Chapter 3 Monitoring Overview 3 10 www zoll com 9650 001355 01 Rev M...
Страница 80: ...Chapter 4 Trends 4 6 www zoll com 9650 001355 01 Rev M...
Страница 90: ...Chapter 5 Alarms 5 10 www zoll com 9650 001355 01 Rev M...
Страница 110: ...Chapter 7 Monitoring Respiration Resp and Heart Rate HR 7 10 www zoll com 9650 001355 01 Rev M...
Страница 126: ...Chapter 8 Monitoring Non Invasive Blood Pressure NIBP 8 16 www zoll com 9650 001355 01 Rev M...
Страница 156: ...Chapter 10 Pulse CO Oximetry SpO2 10 18 www zoll com 9650 001355 01 Rev M...
Страница 166: ...Chapter 11 Monitoring Invasive Pressures IBP 11 10 www zoll com 9650 001355 01 Rev M...
Страница 198: ...Chapter 14 12 Lead ECG Interpretive Analysis 14 16 www zoll com 9650 001355 01 Rev M...
Страница 218: ...Chapter 16 Advisory Defibrillation 16 6 www zoll com 9650 001355 01 Rev M...
Страница 224: ...Chapter 17 Analysis CPR Protocol Defibrillation 17 6 www zoll com 9650 001355 01 Rev M...
Страница 234: ...Chapter 18 External Pacing 18 10 www zoll com 9650 001355 01 Rev M...
Страница 242: ...Chapter 19 Real CPR Help 19 8 www zoll com 9650 001355 01 Rev M...
Страница 254: ...Chapter 21 Patient Data 21 6 www zoll com 9650 001355 01 Rev M...
Страница 338: ...Specifications A 38 www zoll com 9650 001355 01 Rev M...