12-Lead ECG Monitoring Setup
9650-001355-01 Rev. M
X Series Operator’s Guide
14-11
You can then view four 12-Lead snapshot pages by pressing
to step through them
sequentially. For example, the unit displays the following 12-Lead snapshot after the
Analysis
Page 2
:
Fault Conditions Affecting 12-Lead Interpretive Analysis
The X Series unit will not perform Interpretive Analysis if it detects one of the following fault
conditions when attempting to acquire 12-Lead data:
•
A pacer signal is present.
•
A lead fault is detected in the ECG cable.
•
An invalid cable is in use.
If the X Series unit detects one of these fault conditions,
Analysis Page 1
indicates that there is
No data available
for Interpretive Analysis and lists the fault condition; all measurements on
Analysis Page 2
appear as
N/A.
Once you have corrected the fault condition, press
to confirm the correction and perform
the 12-Lead Interpretive Analysis.
12
Содержание X Series
Страница 1: ...Includes Real CPR Help and See Thru CPR X Series Operator s Guide 9650 001355 01 Rev M...
Страница 64: ...Chapter 2 Product Overview 2 24 www zoll com 9650 001355 01 Rev M...
Страница 74: ...Chapter 3 Monitoring Overview 3 10 www zoll com 9650 001355 01 Rev M...
Страница 80: ...Chapter 4 Trends 4 6 www zoll com 9650 001355 01 Rev M...
Страница 90: ...Chapter 5 Alarms 5 10 www zoll com 9650 001355 01 Rev M...
Страница 110: ...Chapter 7 Monitoring Respiration Resp and Heart Rate HR 7 10 www zoll com 9650 001355 01 Rev M...
Страница 126: ...Chapter 8 Monitoring Non Invasive Blood Pressure NIBP 8 16 www zoll com 9650 001355 01 Rev M...
Страница 156: ...Chapter 10 Pulse CO Oximetry SpO2 10 18 www zoll com 9650 001355 01 Rev M...
Страница 166: ...Chapter 11 Monitoring Invasive Pressures IBP 11 10 www zoll com 9650 001355 01 Rev M...
Страница 198: ...Chapter 14 12 Lead ECG Interpretive Analysis 14 16 www zoll com 9650 001355 01 Rev M...
Страница 218: ...Chapter 16 Advisory Defibrillation 16 6 www zoll com 9650 001355 01 Rev M...
Страница 224: ...Chapter 17 Analysis CPR Protocol Defibrillation 17 6 www zoll com 9650 001355 01 Rev M...
Страница 234: ...Chapter 18 External Pacing 18 10 www zoll com 9650 001355 01 Rev M...
Страница 242: ...Chapter 19 Real CPR Help 19 8 www zoll com 9650 001355 01 Rev M...
Страница 254: ...Chapter 21 Patient Data 21 6 www zoll com 9650 001355 01 Rev M...
Страница 338: ...Specifications A 38 www zoll com 9650 001355 01 Rev M...