Specifications
A-36
www.zoll.com
9650-001355-01 Rev. M
ECG Analysis Algorithm Accuracy
Sensitivity and specificity are expressions of ECG analysis algorithm performance when compared to
ECG interpretation by a clinician or expert. Sensitivity refers to the algorithm’s ability to correctly
identify shockable rhythms (as a percentage of the total number of shockable rhythms). Specificity
refers to the algorithm’s ability to correctly identify non-shockable rhythms (as a percentage of the
total number of non-shockable rhythms). The data in the following table summarizes the accuracy of
the ECG analysis algorithm as tested against ZOLL’s ECG Rhythm Database.
The algorithm sequence takes approximately 9 seconds and proceeds as follows:
•
Divides the ECG rhythm into three-second segments.
•
Filters and measures noise, artifact, and baseline wander.
•
Measures baseline content (‘waviness’ at the correct frequencies
—
frequency domain analysis)
of signal.
•
Measures QRS rate, width, and variability.
•
Measures amplitude and temporal regularity (‘auto-correlation’) of peaks and troughs.
•
Determines if multiple 3 second segments are shockable then displays
SHOCK ADVISED
message.
Clinical Performance Results
The performance of the incorporated analysis algorithm in a single analysis sequence satisfies the
applicable requirements specified in ANSI/AAMI DF80 (section 6.8.3) and the recommendations by
Kerber et al. (Circulation. 1997;95(6):1677).
Table A-4. Clinical Performance Results
References:
Young KD, Lewis RJ: “What is confidence? Part 2: Detailed definition and determination of
confidence intervals”. Annals of Emergency Medicine, September 1997; 30; 311-218
William H. Beyer, Ph.D.: “CRC Standard Mathematical Tables 28
th
Edition,” CRC Press, Inc, Boca
Raton, FL., 1981, Percentage Points, F-Distribution Table, pg 573.
Rhythms
Sample
Size
Performance Goals Observed
Performance
90% One-sided
Lower Confidence
Limit
Shockable
Sensitivity
Coarse VF
Rapid VT
536
80
>90%
>75%
>99%
>99%
>99%
>96%
Non-shockable
Specificity
NSR
AF, SB, SVT, Heart block,
idioventricular, PVCs
Asystole
2210
819
115
>99%
>95%
>95%
>99%
>99%
>99%
>99%
>99%
>97%
Intermediate
Fine VF
Other VT
69
28
Report only
Report only
Sensitivity
>89%
>96%
>81%
>84%
Содержание X Series
Страница 1: ...Includes Real CPR Help and See Thru CPR X Series Operator s Guide 9650 001355 01 Rev M...
Страница 64: ...Chapter 2 Product Overview 2 24 www zoll com 9650 001355 01 Rev M...
Страница 74: ...Chapter 3 Monitoring Overview 3 10 www zoll com 9650 001355 01 Rev M...
Страница 80: ...Chapter 4 Trends 4 6 www zoll com 9650 001355 01 Rev M...
Страница 90: ...Chapter 5 Alarms 5 10 www zoll com 9650 001355 01 Rev M...
Страница 110: ...Chapter 7 Monitoring Respiration Resp and Heart Rate HR 7 10 www zoll com 9650 001355 01 Rev M...
Страница 126: ...Chapter 8 Monitoring Non Invasive Blood Pressure NIBP 8 16 www zoll com 9650 001355 01 Rev M...
Страница 156: ...Chapter 10 Pulse CO Oximetry SpO2 10 18 www zoll com 9650 001355 01 Rev M...
Страница 166: ...Chapter 11 Monitoring Invasive Pressures IBP 11 10 www zoll com 9650 001355 01 Rev M...
Страница 198: ...Chapter 14 12 Lead ECG Interpretive Analysis 14 16 www zoll com 9650 001355 01 Rev M...
Страница 218: ...Chapter 16 Advisory Defibrillation 16 6 www zoll com 9650 001355 01 Rev M...
Страница 224: ...Chapter 17 Analysis CPR Protocol Defibrillation 17 6 www zoll com 9650 001355 01 Rev M...
Страница 234: ...Chapter 18 External Pacing 18 10 www zoll com 9650 001355 01 Rev M...
Страница 242: ...Chapter 19 Real CPR Help 19 8 www zoll com 9650 001355 01 Rev M...
Страница 254: ...Chapter 21 Patient Data 21 6 www zoll com 9650 001355 01 Rev M...
Страница 338: ...Specifications A 38 www zoll com 9650 001355 01 Rev M...