Chapter 18 External Pacing
18-4
www.zoll.com
9650-001355-01 Rev. M
4 Set Mode
Use the arrow keys to navigate to Mode, press the Select button, and then use the arrow keys
and the Select button to set the
Pacer Mode to Demand.
Note:
The initial pacing mode can be set in Defib/Pacer Default settings in the Supervisor
Setup menu.
5 Set Pacer Rate
Use the arrow keys to navigate to Rate, press the Select button, and then use the arrow keys and
the Select button to set the
Pacer Rate to a value 10-20 ppm higher than the patient’s intrinsic
heart rate. If no intrinsic rate exists, use 100 ppm. You can increase or decrease the pacer rate
by a value of 5 ppm for rates below 100, and by 10 ppm for rates above 100.
Note:
The initial pacing rate can be set in Defib/Pacer Default settings in the Supervisor
Setup menu.
6 Turn On Pacer
Use the arrow keys to navigate to Start Pacer, then press the Select button to select it. The
Pacing window displays behind the Pacer Settings window.
7 Set Pacer Output
In the Pacer Settings window, use the arrow keys and the Select button to adjust the pacer
output. The pacer output is adjustable in 10 mA increments when increasing the output, and in
5 mA increments when decreasing the output. Observe the ECG for evidence of electrical
capture.
Select the lowest output current that achieves both electrical and mechanical capture.
Note:
If the Pacer Settings window disappears before you have set the output current, press
the
PACER
button again to display the settings window.
PACING
80
ppm
Rate
95
mA
Output
Demand
Mode
Содержание X Series
Страница 1: ...Includes Real CPR Help and See Thru CPR X Series Operator s Guide 9650 001355 01 Rev M...
Страница 64: ...Chapter 2 Product Overview 2 24 www zoll com 9650 001355 01 Rev M...
Страница 74: ...Chapter 3 Monitoring Overview 3 10 www zoll com 9650 001355 01 Rev M...
Страница 80: ...Chapter 4 Trends 4 6 www zoll com 9650 001355 01 Rev M...
Страница 90: ...Chapter 5 Alarms 5 10 www zoll com 9650 001355 01 Rev M...
Страница 110: ...Chapter 7 Monitoring Respiration Resp and Heart Rate HR 7 10 www zoll com 9650 001355 01 Rev M...
Страница 126: ...Chapter 8 Monitoring Non Invasive Blood Pressure NIBP 8 16 www zoll com 9650 001355 01 Rev M...
Страница 156: ...Chapter 10 Pulse CO Oximetry SpO2 10 18 www zoll com 9650 001355 01 Rev M...
Страница 166: ...Chapter 11 Monitoring Invasive Pressures IBP 11 10 www zoll com 9650 001355 01 Rev M...
Страница 198: ...Chapter 14 12 Lead ECG Interpretive Analysis 14 16 www zoll com 9650 001355 01 Rev M...
Страница 218: ...Chapter 16 Advisory Defibrillation 16 6 www zoll com 9650 001355 01 Rev M...
Страница 224: ...Chapter 17 Analysis CPR Protocol Defibrillation 17 6 www zoll com 9650 001355 01 Rev M...
Страница 234: ...Chapter 18 External Pacing 18 10 www zoll com 9650 001355 01 Rev M...
Страница 242: ...Chapter 19 Real CPR Help 19 8 www zoll com 9650 001355 01 Rev M...
Страница 254: ...Chapter 21 Patient Data 21 6 www zoll com 9650 001355 01 Rev M...
Страница 338: ...Specifications A 38 www zoll com 9650 001355 01 Rev M...