9650-001355-01 Rev. M
X Series Operator’s Guide
9-1
Chapter 9
Monitoring CO
2
This chapter describes how to use the X Series unit to monitor End Tidal Carbon Dioxide
(EtCO
2
), breath rate, and Fractional Inspired Carbon Dioxide (FiCO
2
). These options use the
same connector on the X Series unit and may be used interchangeably.
Overview
The X Series unit uses the Oridion Microstream FilterLine
®
and Smart CapnoLine
®
sidestream
systems to monitor CO
2
.
The FilterLine and Smart CapnoLine sidestream systems draw small samples of gas from the
patient’s airway via a nasal/oral cannula or airway adapter, and pass these gases through a solid
state infrared sensor (located away from the patient’s airway) that measures CO
2
. The
Microstream system can be used for CO
2
measurements on intubated and non-intubated infant,
pediatric, and adult patients.
The Microstream CO
2
sensor generates infrared light, and beams it through the sampled
breathing gases and determines CO
2
concentration by measuring the amount of light absorbed
by gases.
The X Series
unit displays EtCO
2
(the concentration of carbon dioxide detected at the end of
each exhalation) and FiCO as a numerical value in millimeters of mercury (mmHg). In
addition, the unit can display a capnogram. The capnogram is a valuable clinical tool that can
be used to assess patient airway integrity and proper endotracheal (ET) tube placement. The
unit calculates respiration rate by measuring the time interval between detected peaks of the
CO
2
waveform. The technology differentiates between waveforms caused by breathing and
those caused by cardiogenic oscillations and artifact.
The X Series is equipped with automatic barometric pressure compensation.
Содержание X Series
Страница 1: ...Includes Real CPR Help and See Thru CPR X Series Operator s Guide 9650 001355 01 Rev M...
Страница 64: ...Chapter 2 Product Overview 2 24 www zoll com 9650 001355 01 Rev M...
Страница 74: ...Chapter 3 Monitoring Overview 3 10 www zoll com 9650 001355 01 Rev M...
Страница 80: ...Chapter 4 Trends 4 6 www zoll com 9650 001355 01 Rev M...
Страница 90: ...Chapter 5 Alarms 5 10 www zoll com 9650 001355 01 Rev M...
Страница 110: ...Chapter 7 Monitoring Respiration Resp and Heart Rate HR 7 10 www zoll com 9650 001355 01 Rev M...
Страница 126: ...Chapter 8 Monitoring Non Invasive Blood Pressure NIBP 8 16 www zoll com 9650 001355 01 Rev M...
Страница 156: ...Chapter 10 Pulse CO Oximetry SpO2 10 18 www zoll com 9650 001355 01 Rev M...
Страница 166: ...Chapter 11 Monitoring Invasive Pressures IBP 11 10 www zoll com 9650 001355 01 Rev M...
Страница 198: ...Chapter 14 12 Lead ECG Interpretive Analysis 14 16 www zoll com 9650 001355 01 Rev M...
Страница 218: ...Chapter 16 Advisory Defibrillation 16 6 www zoll com 9650 001355 01 Rev M...
Страница 224: ...Chapter 17 Analysis CPR Protocol Defibrillation 17 6 www zoll com 9650 001355 01 Rev M...
Страница 234: ...Chapter 18 External Pacing 18 10 www zoll com 9650 001355 01 Rev M...
Страница 242: ...Chapter 19 Real CPR Help 19 8 www zoll com 9650 001355 01 Rev M...
Страница 254: ...Chapter 21 Patient Data 21 6 www zoll com 9650 001355 01 Rev M...
Страница 338: ...Specifications A 38 www zoll com 9650 001355 01 Rev M...