NIBP System Messages
9650-001355-01 Rev. M
X Series Operator’s Guide
8-15
NIBP System Messages
When monitoring NIBP, the X Series unit may display the following messages:
System Message
Cause
READING IN PROGRESS
The unit is taking an NIBP measurement and
functioning normally.
READING STOPPED
The unit has stopped an NIBP measurement,
because the operator has pressed the NIBP button
and cancelled the measurement.
READING FAILED
The patient’s pulse is too weak to obtain an NIBP
measurement, or the cuff requires adjustment.
ARTIFACT
Excessive artifact is preventing the NIBP
measurement. If ECG is available, consider turning
SmartCuf on.
NEONATE ALERT
The unit has detected a neonate cuff when in Adult
mode. Replace cuff or correct patient mode, as
appropriate. If the detection is in error, dismiss the
alert and reattempt the NIBP measurement.
AIR LEAK
A major air leak is preventing cuff inflation. Check
hose and cuff connections, replace a defective hose
or cuff, as necessary, and reattempt NIBP
measurement.
KINKED HOSE
An air obstruction is preventing the correct operation
of the unit. Check hose for kinks. If using a dual
lumen hose, verify the correct placement of the O
ring on the hose connector. Eliminate kinks or
replace hose, if necessary.
NIBP DISABLED
A system error has occurred, and the X Series unit
should be serviced.
NIBP CALIBRATION CHECK DUE
The annual calibration check is due. You can
continue to use the device, however, we recommend
that you contact your trained specialist or the ZOLL
Technical Service Department or a ZOLL authorized
service representative as soon as possible to
perform the recommended annual NIBP calibration
check.
PLEASE WAIT
After completing an NIBP request in Automatic mode
or STAT (TurboCuf) mode, the unit must wait up to
30 seconds before initiating the next NIBP
measurement request.
PRESS. LIMIT EXCEEDED
The cuff pressure has exceeded the allowable limit.
Restrict patient movement and check the hose for
kinks. Eliminate kinks or replace the hose, if
necessary.
TEMP OUT OF RANGE
The NIBP module is outside of its operating
temperature range. Move unit to proper operating
temperature and reattempt the NIBP measurement.
WEAK PULSE
The patient’s pulse is too weak to obtain an NIBP
measurement or the cuff requires adjustment.
Содержание X Series
Страница 1: ...Includes Real CPR Help and See Thru CPR X Series Operator s Guide 9650 001355 01 Rev M...
Страница 64: ...Chapter 2 Product Overview 2 24 www zoll com 9650 001355 01 Rev M...
Страница 74: ...Chapter 3 Monitoring Overview 3 10 www zoll com 9650 001355 01 Rev M...
Страница 80: ...Chapter 4 Trends 4 6 www zoll com 9650 001355 01 Rev M...
Страница 90: ...Chapter 5 Alarms 5 10 www zoll com 9650 001355 01 Rev M...
Страница 110: ...Chapter 7 Monitoring Respiration Resp and Heart Rate HR 7 10 www zoll com 9650 001355 01 Rev M...
Страница 126: ...Chapter 8 Monitoring Non Invasive Blood Pressure NIBP 8 16 www zoll com 9650 001355 01 Rev M...
Страница 156: ...Chapter 10 Pulse CO Oximetry SpO2 10 18 www zoll com 9650 001355 01 Rev M...
Страница 166: ...Chapter 11 Monitoring Invasive Pressures IBP 11 10 www zoll com 9650 001355 01 Rev M...
Страница 198: ...Chapter 14 12 Lead ECG Interpretive Analysis 14 16 www zoll com 9650 001355 01 Rev M...
Страница 218: ...Chapter 16 Advisory Defibrillation 16 6 www zoll com 9650 001355 01 Rev M...
Страница 224: ...Chapter 17 Analysis CPR Protocol Defibrillation 17 6 www zoll com 9650 001355 01 Rev M...
Страница 234: ...Chapter 18 External Pacing 18 10 www zoll com 9650 001355 01 Rev M...
Страница 242: ...Chapter 19 Real CPR Help 19 8 www zoll com 9650 001355 01 Rev M...
Страница 254: ...Chapter 21 Patient Data 21 6 www zoll com 9650 001355 01 Rev M...
Страница 338: ...Specifications A 38 www zoll com 9650 001355 01 Rev M...