Configuring NIBP Alarms and Settings
9650-001355-01 Rev. M
X Series Operator’s Guide
8-11
Configuring NIBP Alarms and Settings
The last step in preparing to perform NIBP measurements is to ensure that the necessary alarms
are enabled (or disabled), that alarm limits are appropriate, and that the NIBP settings are
correct.
Enabling/Disabling NIBP Alarms and Setting Alarm Limits
When enabled, the X Series unit sounds alarms whenever measurements are outside set limits
for the following:
•
High and Low Systolic Pressure
•
High and Low Diastolic Pressure
•
High and Low Mean Arterial Pressure (MAP)
To enable (or disable) NIBP alarms and set upper and lower alarm limits, you can do so through
the Alarms
quick access key (
) or the
NIBP Parameter Control Panel
.
To configure NIBP alarm limits through the Alarms quick access key:
1. Press the More quick access key (
).
2.
Press the Alarms quick access key (
).
3.
Press the
Limits
quick access key (
).
4.
Use the navigation keys to highlight and select the appropriate alarm menu selection. For
NIBP, the alarm menu selections are
NIBP Systolic Alarm, NIBP Diastolic Alarm,
and
NIBP MAP Alarm
.
5.
On the selected
NIBP alarm settings menu, use the navigation keys to select the fields that
you want to change. The fields are
•
Upper Limit Enable
•
Lower Limit Enable
•
Upper Limit
•
Lower Limit
6.
When you are finished changing values on the alarm
settings menu, select the back arrow to
exit the menu.
Setting Upper and Lower Systolic Alarm Limits
Initially, the
NIBP Systolic Alarm Settings
menu specifies that the NIBP systolic pressure
alarms are enabled (
ON
)
or disabled (
OFF
),
and displays the default upper and lower systolic
limits. The upper and lower limits can be
ON
or
OFF
(default is
OFF
). The following table lists
the default NIBP Systolic limits for adult, pediatric, and neonate patients, and gives the range in
which you can set these limits:
Patient Type
NIBP Systolic Limit Default
NIBP Systolic Limit Range
Adult
Lower: 75 mmHg
Upper: 220 mmHg
Lower: 30-258 mmHg
Upper: 32-260 mmHg
Pediatric
Lower: 75 mmHg
Upper: 145 mmHg
Lower: 30-158 mmHg
Upper: 32-160 mmHg
Neonate
Lower: 50 mmHg
Upper: 100 mmHg
Lower: 20-118 mmHg
Upper: 22-120 mmHg
Limits
Содержание X Series
Страница 1: ...Includes Real CPR Help and See Thru CPR X Series Operator s Guide 9650 001355 01 Rev M...
Страница 64: ...Chapter 2 Product Overview 2 24 www zoll com 9650 001355 01 Rev M...
Страница 74: ...Chapter 3 Monitoring Overview 3 10 www zoll com 9650 001355 01 Rev M...
Страница 80: ...Chapter 4 Trends 4 6 www zoll com 9650 001355 01 Rev M...
Страница 90: ...Chapter 5 Alarms 5 10 www zoll com 9650 001355 01 Rev M...
Страница 110: ...Chapter 7 Monitoring Respiration Resp and Heart Rate HR 7 10 www zoll com 9650 001355 01 Rev M...
Страница 126: ...Chapter 8 Monitoring Non Invasive Blood Pressure NIBP 8 16 www zoll com 9650 001355 01 Rev M...
Страница 156: ...Chapter 10 Pulse CO Oximetry SpO2 10 18 www zoll com 9650 001355 01 Rev M...
Страница 166: ...Chapter 11 Monitoring Invasive Pressures IBP 11 10 www zoll com 9650 001355 01 Rev M...
Страница 198: ...Chapter 14 12 Lead ECG Interpretive Analysis 14 16 www zoll com 9650 001355 01 Rev M...
Страница 218: ...Chapter 16 Advisory Defibrillation 16 6 www zoll com 9650 001355 01 Rev M...
Страница 224: ...Chapter 17 Analysis CPR Protocol Defibrillation 17 6 www zoll com 9650 001355 01 Rev M...
Страница 234: ...Chapter 18 External Pacing 18 10 www zoll com 9650 001355 01 Rev M...
Страница 242: ...Chapter 19 Real CPR Help 19 8 www zoll com 9650 001355 01 Rev M...
Страница 254: ...Chapter 21 Patient Data 21 6 www zoll com 9650 001355 01 Rev M...
Страница 338: ...Specifications A 38 www zoll com 9650 001355 01 Rev M...