ECG Monitoring Setup
9650-001355-01 Rev. M
X Series Operator’s Guide
6-7
The X Series unit is configured to display
Pads
by default for the top trace. If configured, the
unit can also display four ECG traces on startup, when no other monitoring devices are
attached.
The are two ways to specify which ECG lead is the source of the primary waveform trace. One
way is to press the ECG lead selection quick access key
to display the available ECG
waveform sources. The available waveform sources are determined by the type of ECG cable
connected to the unit.
The other way to specify the source of the primary waveform trace is to navigate to and select
the source label for the primary ECG waveform (Lead
l
in the screen below). The X Series unit
then displays the available ECG waveform sources. The following example illustrates the
waveform source list that the X Series unit displays when a 5-lead ECG cable is connected to it.
The list of available ECG waveform sources includes Leads
I
,
II
,
III
,
aVR
,
aVL
,
aVF
, and
V
. You
can select Lead
lI
(the default), or use the navigation keys to highlight and select another
displayed ECG lead as the source for the waveform trace.
Select a current waveform source, and the unit displays the waveform immediately. If you
select a waveform source that is not currently available, the unit displays the message,
LEAD
FAULT
.
For more information on how to configure the display of waveforms on the X Series unit, see
Chapter 3,
Monitoring Overview
.
I, II,
III...
06/06/2011 12:34:56
Adult
00:17:43
I, II,
III...
12
CO
2
R
SYNC
1 cm/mV
I
HR
bpm
80
121
79
(96)
NIBP
mmHg
38
12
BR
CO2
mmHg
SpO2
%
97
T1
ºF
98.6
Pads
III
aVL
aVF
V
Insert
Source
II
I
Cascade
aVR
Содержание X Series
Страница 1: ...Includes Real CPR Help and See Thru CPR X Series Operator s Guide 9650 001355 01 Rev M...
Страница 64: ...Chapter 2 Product Overview 2 24 www zoll com 9650 001355 01 Rev M...
Страница 74: ...Chapter 3 Monitoring Overview 3 10 www zoll com 9650 001355 01 Rev M...
Страница 80: ...Chapter 4 Trends 4 6 www zoll com 9650 001355 01 Rev M...
Страница 90: ...Chapter 5 Alarms 5 10 www zoll com 9650 001355 01 Rev M...
Страница 110: ...Chapter 7 Monitoring Respiration Resp and Heart Rate HR 7 10 www zoll com 9650 001355 01 Rev M...
Страница 126: ...Chapter 8 Monitoring Non Invasive Blood Pressure NIBP 8 16 www zoll com 9650 001355 01 Rev M...
Страница 156: ...Chapter 10 Pulse CO Oximetry SpO2 10 18 www zoll com 9650 001355 01 Rev M...
Страница 166: ...Chapter 11 Monitoring Invasive Pressures IBP 11 10 www zoll com 9650 001355 01 Rev M...
Страница 198: ...Chapter 14 12 Lead ECG Interpretive Analysis 14 16 www zoll com 9650 001355 01 Rev M...
Страница 218: ...Chapter 16 Advisory Defibrillation 16 6 www zoll com 9650 001355 01 Rev M...
Страница 224: ...Chapter 17 Analysis CPR Protocol Defibrillation 17 6 www zoll com 9650 001355 01 Rev M...
Страница 234: ...Chapter 18 External Pacing 18 10 www zoll com 9650 001355 01 Rev M...
Страница 242: ...Chapter 19 Real CPR Help 19 8 www zoll com 9650 001355 01 Rev M...
Страница 254: ...Chapter 21 Patient Data 21 6 www zoll com 9650 001355 01 Rev M...
Страница 338: ...Specifications A 38 www zoll com 9650 001355 01 Rev M...