![ZOLL X Series Скачать руководство пользователя страница 321](http://html1.mh-extra.com/html/zoll/x-series/x-series_operators-manual_4055648321.webp)
Pulse Oximeter
9650-001355-01 Rev. M
X Series Operator’s Guide
A-21
Alarm Limits:
On/Off displayed on monitor. User selectable.
SpO
2
: High 72 - 100% saturation, Low 70 - 98% saturation
SpCO: High 2 - 100% saturation, Low 0 - 99% saturation
SpMet: High 1 - 100% saturation, Low 0 - 99% saturation
SpHb: High 2 - 25 g/dL, Low 0 - 24.9 g/dL
SpOC: High 0.1 – 35 ml/dL, Low 0 – 34.9 ml/dL
PVI: High 2 – 100%, Low 0 – 98%
PI: High 0.2 – 20%, Low 0 – 19.8%
Pulse Rate: High 60 - 235 beats per minute, Low 20 - 100 beats per minute
SpO
2
Wavelength for Sensors:
Masimo's LNOP and LNCS sensors use Red and Infrared light emitting diodes.
The wavelengths for all of the sensors except LNOP tip clip (LNOP TC-I),
LNCS/M-LNCS tip clip (LNCS/M-LNCS TC-I), LNOP transflectance (LNOP
TF-1), and LNCS/M- LNCS transflectance (LNCS/M-LNCS TF-1), are
identified as follows:
The LNOP tip clip (LNOP TC-I) and LNCS/M-LNCS tip clip (LNCS/M-
LNCS TC-I) sensors use different light emitting diodes. The wavelength
information is as follows:
T h e L N O P transflectance (LNCS/M-LNCS
TF-I)
forehead sensors use
different light emitting diodes. The wavelength information is as follows:
For Sp02 calculations with a rainbow sensor, the wavelength values shown
inthe above tables are the same. For rainbow parameter measurements,
sensors use light emitting diodes in both the visible and infrared spectrum from
the 500 nm to 1400 nm range.
Energies (Radiant Power) of light for
LNCS Sensors at 50 mA pulsed:
15 mW
Energies (Radiant Power) of light for
Rainbow Sensors at 100 mA pulsed:
25 mW
Bio-Compatibility:
Patient contacting material meets requirements of ISO 10993-1, Biological
Evaluation of Medical Device - Part I, for external devices, intact surfaces and
short-term exposure
LED
Wavelength
Red
660 nm
Infrared
905 nm
LED
Wavelength
Red
653 nm
Infrared
880 nm
LED
Wavelength
Red
660 nm
Infrared
880 nm
Содержание X Series
Страница 1: ...Includes Real CPR Help and See Thru CPR X Series Operator s Guide 9650 001355 01 Rev M...
Страница 64: ...Chapter 2 Product Overview 2 24 www zoll com 9650 001355 01 Rev M...
Страница 74: ...Chapter 3 Monitoring Overview 3 10 www zoll com 9650 001355 01 Rev M...
Страница 80: ...Chapter 4 Trends 4 6 www zoll com 9650 001355 01 Rev M...
Страница 90: ...Chapter 5 Alarms 5 10 www zoll com 9650 001355 01 Rev M...
Страница 110: ...Chapter 7 Monitoring Respiration Resp and Heart Rate HR 7 10 www zoll com 9650 001355 01 Rev M...
Страница 126: ...Chapter 8 Monitoring Non Invasive Blood Pressure NIBP 8 16 www zoll com 9650 001355 01 Rev M...
Страница 156: ...Chapter 10 Pulse CO Oximetry SpO2 10 18 www zoll com 9650 001355 01 Rev M...
Страница 166: ...Chapter 11 Monitoring Invasive Pressures IBP 11 10 www zoll com 9650 001355 01 Rev M...
Страница 198: ...Chapter 14 12 Lead ECG Interpretive Analysis 14 16 www zoll com 9650 001355 01 Rev M...
Страница 218: ...Chapter 16 Advisory Defibrillation 16 6 www zoll com 9650 001355 01 Rev M...
Страница 224: ...Chapter 17 Analysis CPR Protocol Defibrillation 17 6 www zoll com 9650 001355 01 Rev M...
Страница 234: ...Chapter 18 External Pacing 18 10 www zoll com 9650 001355 01 Rev M...
Страница 242: ...Chapter 19 Real CPR Help 19 8 www zoll com 9650 001355 01 Rev M...
Страница 254: ...Chapter 21 Patient Data 21 6 www zoll com 9650 001355 01 Rev M...
Страница 338: ...Specifications A 38 www zoll com 9650 001355 01 Rev M...