![background image](http://html1.mh-extra.com/html/zoll/x-series/x-series_operators-manual_4055648152.webp)
Chapter 10 Pulse CO-Oximetry (SpO2)
10-14
www.zoll.com
9650-001355-01 Rev. M
Using the SpO
2
Parameter Control Panel
Use the navigation keys to highlight and select the SpO
2
numeric display to display the SpO2
Parameter Control Panel
,
where you can select options that optimize the use of SpO
2
for the
patient:
SpCO and SpMet, and SpHb, SpOC, PVI and PI Monitoring
If the SpCO and SpMet or the SpHb, SpOC, PVI and PI options are installed on your unit you
can enable or disable monitoring these parameters on the SpO2 Control Panel.
Specifying the SpO
2
Averaging Time
The Masimo SpO
2
module in the X Series unit provides three different time periods over which
SpO
2
values are averaged: 4 seconds, 8 seconds (default) and 16 seconds. The averaging period
is rarely changed from the 8 second default setting. For high risk patients with rapidly changing
SpO
2
conditions, use the 4 second setting. Use the 16 second setting only when the 8 second
setting (default) is inadequate due to extremely high artifact conditions.
Pediatric
Lower: 0%
Upper: 20%
Lower: 0
-
98%
Upper: 0.2 - 20%
Neonate
Lower: 0%
Upper: 20%
Lower: 0
-
98%
Upper: 0.2 - 20%
Patient Type
PI Limit Default
PI Limit Range
Содержание X Series
Страница 1: ...Includes Real CPR Help and See Thru CPR X Series Operator s Guide 9650 001355 01 Rev M...
Страница 64: ...Chapter 2 Product Overview 2 24 www zoll com 9650 001355 01 Rev M...
Страница 74: ...Chapter 3 Monitoring Overview 3 10 www zoll com 9650 001355 01 Rev M...
Страница 80: ...Chapter 4 Trends 4 6 www zoll com 9650 001355 01 Rev M...
Страница 90: ...Chapter 5 Alarms 5 10 www zoll com 9650 001355 01 Rev M...
Страница 110: ...Chapter 7 Monitoring Respiration Resp and Heart Rate HR 7 10 www zoll com 9650 001355 01 Rev M...
Страница 126: ...Chapter 8 Monitoring Non Invasive Blood Pressure NIBP 8 16 www zoll com 9650 001355 01 Rev M...
Страница 156: ...Chapter 10 Pulse CO Oximetry SpO2 10 18 www zoll com 9650 001355 01 Rev M...
Страница 166: ...Chapter 11 Monitoring Invasive Pressures IBP 11 10 www zoll com 9650 001355 01 Rev M...
Страница 198: ...Chapter 14 12 Lead ECG Interpretive Analysis 14 16 www zoll com 9650 001355 01 Rev M...
Страница 218: ...Chapter 16 Advisory Defibrillation 16 6 www zoll com 9650 001355 01 Rev M...
Страница 224: ...Chapter 17 Analysis CPR Protocol Defibrillation 17 6 www zoll com 9650 001355 01 Rev M...
Страница 234: ...Chapter 18 External Pacing 18 10 www zoll com 9650 001355 01 Rev M...
Страница 242: ...Chapter 19 Real CPR Help 19 8 www zoll com 9650 001355 01 Rev M...
Страница 254: ...Chapter 21 Patient Data 21 6 www zoll com 9650 001355 01 Rev M...
Страница 338: ...Specifications A 38 www zoll com 9650 001355 01 Rev M...