Specifications
A-2
www.zoll.com
9650-001355-01 Rev. M
Defibrillator
Charge Time
:
•
Less than 7 seconds with a new, fully charged battery (first 15 charges to 200 joules).
•
For the sixteenth discharge at maximum energy, the charge time is less than 10 seconds.
Depleted batteries result in a longer defibrillator charge time.
•
Less than 15 seconds when operating without a battery, using AC power alone at 90% of the
rated mains voltage.
•
Less than 25 seconds from the initial power on, with a new, fully charged battery pack
(depleted by up to fifteen 200 joule discharges) or when operating without a battery, using
AC power alone at 90% of the rated mains voltage.
Rhythm Analysis and Charge Time in AED Mode
•
Less than 30 seconds with a new, fully charged battery (first 15 charges to 200 joules).
•
For the sixteenth discharge at maximum energy, the analysis and charge time is less than 30
seconds. Depleted batteries result in a longer defibrillator charge time.
•
Less than 30 seconds when operating without a battery, using AC power alone at 90% of the
rated mains voltage.
•
Less than 40 seconds from the initial power on, with a new, fully charged battery pack
(depleted by up to fifteen 200 joule discharges) or when operating without a battery, using
AC power alone at 90% of the rated mains voltage.
Patient Impedance Range:
10- 300 ohms
Synchronized Mode
: Synchronizes defibrillator discharge to the patient’s R wave. SYNC is
indicated on the display with R wave markers above the ECG waveform on the screen and
stripchart. When ECG is monitored by the device, meets the DF-80:2003 requirement of 60ms
maximum time delay between the peak of the R wave and the delivery of energy.
Table A-1 shows the characteristics of the X Series Rectilinear Biphasic™ waveform when
discharged into 25 ohm, 50 ohm, 100 ohm, 125 ohm, 150 ohm and 175 ohm loads at the
maximum energy setting of 200 joules.
Table A-1. X Series Rectilinear Biphasic Waveform Characteristics
200 J discharged into
25
50
100
125
150
175
First phase
Maximum initial current
31.4 A
30.4 A
19.7 A
19.4 A
16.7 A
15.6 A
Average current
27.1 A
24.9 A
17.5 A
16.2 A
14.4 A
13.2 A
Duration
6 ms
6 ms
6 ms
6 ms
6 ms
6 ms
Interphase duration
(between first and
second phases)
200
s
200
s
200
s
200
s
200
s
200
s
Second phase
Initial current
29.2 A
18.8 A
15.1 A
13.2 A
12.1 A
11 A
Average current
14.7 A
13 A
12.5 A
11.3 A
10.7 A
9.9 A
Duration
4 ms
4 ms
4 ms
4 ms
4 ms
4 ms
Содержание X Series
Страница 1: ...Includes Real CPR Help and See Thru CPR X Series Operator s Guide 9650 001355 01 Rev M...
Страница 64: ...Chapter 2 Product Overview 2 24 www zoll com 9650 001355 01 Rev M...
Страница 74: ...Chapter 3 Monitoring Overview 3 10 www zoll com 9650 001355 01 Rev M...
Страница 80: ...Chapter 4 Trends 4 6 www zoll com 9650 001355 01 Rev M...
Страница 90: ...Chapter 5 Alarms 5 10 www zoll com 9650 001355 01 Rev M...
Страница 110: ...Chapter 7 Monitoring Respiration Resp and Heart Rate HR 7 10 www zoll com 9650 001355 01 Rev M...
Страница 126: ...Chapter 8 Monitoring Non Invasive Blood Pressure NIBP 8 16 www zoll com 9650 001355 01 Rev M...
Страница 156: ...Chapter 10 Pulse CO Oximetry SpO2 10 18 www zoll com 9650 001355 01 Rev M...
Страница 166: ...Chapter 11 Monitoring Invasive Pressures IBP 11 10 www zoll com 9650 001355 01 Rev M...
Страница 198: ...Chapter 14 12 Lead ECG Interpretive Analysis 14 16 www zoll com 9650 001355 01 Rev M...
Страница 218: ...Chapter 16 Advisory Defibrillation 16 6 www zoll com 9650 001355 01 Rev M...
Страница 224: ...Chapter 17 Analysis CPR Protocol Defibrillation 17 6 www zoll com 9650 001355 01 Rev M...
Страница 234: ...Chapter 18 External Pacing 18 10 www zoll com 9650 001355 01 Rev M...
Страница 242: ...Chapter 19 Real CPR Help 19 8 www zoll com 9650 001355 01 Rev M...
Страница 254: ...Chapter 21 Patient Data 21 6 www zoll com 9650 001355 01 Rev M...
Страница 338: ...Specifications A 38 www zoll com 9650 001355 01 Rev M...