Chapter 11 Monitoring Invasive Pressures (IBP)
11-4
www.zoll.com
9650-001355-01 Rev. M
Zeroing the Transducer
To ensure that the X Series unit measures pressure accurately, you must zero the transducer
before each use. If you change or disconnect a transducer, you must zero the new transducer
before use. If you move a transducer to a different monitor, you must zero the transducer again,
even if you have already done so on another unit. In addition to the procedure below, follow the
transducer manufacturer’s
Directions for Use
and your organization’s established clinical
protocol.
Follow these steps when zeroing a transducer:
1. Place the transducer at the same height as the patient’s left atrium.
2.
Close the transducer stopcock to the patient.
3.
Open the transducer’s venting stopcock to atmospheric air.
4.
Allow the transducer a few seconds to settle.
5.
Use the navigation keys to highlight and select the tranducer’s IBP channel and display the
IBP channel’s parameter control panel:
Figure 11-2 IBP Channel Parameter Control Panel
6.
Select
Zero Probe
. The unit displays the message
ZEROING
in the IBP Channel’s numeric
display.
7.
The message
ZEROED
appears in the IBP channel’s numeric display.
8.
Close the transducer’s stopcock.
9.
If the unit was unable to zero the transducer, the message
ZERO REJECTED
appears in the
IBP channel’s numeric display. The unit will not display pressure values for the IBP channel
until it has successfully zeroed the transducer and established an acceptable zero reference.
Check that the unit is open to atmospheric air and that it is properly connected to the unit, then
try zeroing the transducer again. The X Series unit will not zero the transducer if it detects
pulsation in the pressure channel, if there is too much noise in the signal, or if transducer’s
offset is too great.
If you cannot zero the transducer after several attempts, replace the transducer or the transducer
cable.
P1
Lower
Upper
P1 Systolic Alarm
90
150
Source Label
P1
Display Format
S/D (M)
P1 Diastolic Alarm
60
90
P1 Mean Alarm
70
120
Zero Probe
Содержание X Series
Страница 1: ...Includes Real CPR Help and See Thru CPR X Series Operator s Guide 9650 001355 01 Rev M...
Страница 64: ...Chapter 2 Product Overview 2 24 www zoll com 9650 001355 01 Rev M...
Страница 74: ...Chapter 3 Monitoring Overview 3 10 www zoll com 9650 001355 01 Rev M...
Страница 80: ...Chapter 4 Trends 4 6 www zoll com 9650 001355 01 Rev M...
Страница 90: ...Chapter 5 Alarms 5 10 www zoll com 9650 001355 01 Rev M...
Страница 110: ...Chapter 7 Monitoring Respiration Resp and Heart Rate HR 7 10 www zoll com 9650 001355 01 Rev M...
Страница 126: ...Chapter 8 Monitoring Non Invasive Blood Pressure NIBP 8 16 www zoll com 9650 001355 01 Rev M...
Страница 156: ...Chapter 10 Pulse CO Oximetry SpO2 10 18 www zoll com 9650 001355 01 Rev M...
Страница 166: ...Chapter 11 Monitoring Invasive Pressures IBP 11 10 www zoll com 9650 001355 01 Rev M...
Страница 198: ...Chapter 14 12 Lead ECG Interpretive Analysis 14 16 www zoll com 9650 001355 01 Rev M...
Страница 218: ...Chapter 16 Advisory Defibrillation 16 6 www zoll com 9650 001355 01 Rev M...
Страница 224: ...Chapter 17 Analysis CPR Protocol Defibrillation 17 6 www zoll com 9650 001355 01 Rev M...
Страница 234: ...Chapter 18 External Pacing 18 10 www zoll com 9650 001355 01 Rev M...
Страница 242: ...Chapter 19 Real CPR Help 19 8 www zoll com 9650 001355 01 Rev M...
Страница 254: ...Chapter 21 Patient Data 21 6 www zoll com 9650 001355 01 Rev M...
Страница 338: ...Specifications A 38 www zoll com 9650 001355 01 Rev M...