Chapter 1 General Information
1-24
www.zoll.com
9650-001355-01 Rev. M
Cautions
•
If the unit is to be stored longer than 30 days, remove the battery pack.
•
Do not sterilize the defibrillator, or its accessories unless the accessories are labeled as
sterilizable.
•
Do not immerse any part of the defibrillator in water.
•
Do not use the defibrillator if excessive condensation is visible on the device. Wipe only the
outside with a damp cloth.
•
Do not use ketones (such as acetone or MEK) on the defibrillator.
•
Avoid using abrasives (including paper towels) on the display window.
•
To achieve the specified level of protection against spilled or splashed liquids, thoroughly
dry all exposed surfaces of this device prior to operation or connections to auxiliary power.
•
If liquids enter the device connectors, remove all liquid from the connectors and allow the
device to dry thoroughly prior to use.
•
Grounding reliability can be achieved only when the equipment is connected to a receptacle
marked “HOSPITAL ONLY,” “HOSPITAL GRADE,” or equivalent. If the grounding
integrity of the line cord or ac receptacle is questionable, operate the defibrillator using
battery power only.
•
Do not connect to an electrical outlet controlled by a wall switch or dimmer.
•
To protect the unit from damage during defibrillation, for accurate ECG information, and to
protect against noise and other interference, use only internal current-limiting ECG cables
specified or supplied by ZOLL.
•
For continued safety and EMI performance, use only the line cord supplied by ZOLL.
•
Electrical installation of the room or the building in which the monitor is to be used must
comply with regulations specified by the country in which the equipment is to be used.
•
Dispose of battery packs in accordance with national, regional and local regulations. Battery
packs should be shipped to a reclamation facility for recovery of metal and plastic
compounds as the proper method of waste management.
•
Do not place the device where the controls can be changed by the patient.
Restarting the Defibrillator
Certain events require the X Series products to be restarted after they shut off or become
inoperative (for example, when the battery runs down and the unit shuts off).
In such a case, always try to restore defibrillator operation as follows:
1. Press the power switch on the top of the unit to turn it off.
2.
If necessary, replace a depleted battery with a fully charged pack, or connect the defibrillator
to auxiliary power.
3.
Press the power switch on the top of the unit to turn it back on.
This sequence is necessary to restart the defibrillator and can also be used to clear some fault
messages when immediate use of the defibrillator is required.
If the X Series unit is powered off for less than 2 minutes, all patient monitoring parameter
settings will be retained. If the unit has been powered off for at least two minutes, it will be
considered a New Patient and all of the patient-specific parameters (alarm limits, defibrillator
energy, etc.) will be reset to their default values.
Содержание X Series
Страница 1: ...Includes Real CPR Help and See Thru CPR X Series Operator s Guide 9650 001355 01 Rev M...
Страница 64: ...Chapter 2 Product Overview 2 24 www zoll com 9650 001355 01 Rev M...
Страница 74: ...Chapter 3 Monitoring Overview 3 10 www zoll com 9650 001355 01 Rev M...
Страница 80: ...Chapter 4 Trends 4 6 www zoll com 9650 001355 01 Rev M...
Страница 90: ...Chapter 5 Alarms 5 10 www zoll com 9650 001355 01 Rev M...
Страница 110: ...Chapter 7 Monitoring Respiration Resp and Heart Rate HR 7 10 www zoll com 9650 001355 01 Rev M...
Страница 126: ...Chapter 8 Monitoring Non Invasive Blood Pressure NIBP 8 16 www zoll com 9650 001355 01 Rev M...
Страница 156: ...Chapter 10 Pulse CO Oximetry SpO2 10 18 www zoll com 9650 001355 01 Rev M...
Страница 166: ...Chapter 11 Monitoring Invasive Pressures IBP 11 10 www zoll com 9650 001355 01 Rev M...
Страница 198: ...Chapter 14 12 Lead ECG Interpretive Analysis 14 16 www zoll com 9650 001355 01 Rev M...
Страница 218: ...Chapter 16 Advisory Defibrillation 16 6 www zoll com 9650 001355 01 Rev M...
Страница 224: ...Chapter 17 Analysis CPR Protocol Defibrillation 17 6 www zoll com 9650 001355 01 Rev M...
Страница 234: ...Chapter 18 External Pacing 18 10 www zoll com 9650 001355 01 Rev M...
Страница 242: ...Chapter 19 Real CPR Help 19 8 www zoll com 9650 001355 01 Rev M...
Страница 254: ...Chapter 21 Patient Data 21 6 www zoll com 9650 001355 01 Rev M...
Страница 338: ...Specifications A 38 www zoll com 9650 001355 01 Rev M...