Lamp replacement and calibration - Corona process | Service manual |
EN_40_070_0010I
Sensor calibration
Lamp replacement and calibration Corona process
Note
The program sequence must be fully completed for both the default
lamp and the spare lamp.
4
Resetting the lamp counter
Reset the lamp lifetime counter by clicking
Yes
.
Fig. 32 Resetting the lamp counter
5
Selecting the certificate of the white standard
Note
Ensure that the white standard certification is still valid.
The corresponding certificate can be found on the CD included in the delivery
package. If no CD drive is available on the test computer the certificate must
be copied to another storage device. The certificate is mandatory.
Compare the report number on the white standard with the report number
on the CD and select the appropriate csv file.