Service manual | Lamp replacement and calibration - Corona process
Replacing halogen lamps
Lamp replacement and calibration Corona process
4.4 Removing halogen lamps – lamp type B
1
Disconnecting the lamp cable
Lamp L1 / Default lamp:
ȩ
Unplug the lamp cable
4
from the board (see Fig. 22).
ȩ
Open the cable holder
1
and remove the cable connectors there.
ȩ
Unplug the lamp cable.
Lamp L2 / Spare lamp:
ȩ
Unplug the lamp cable
3
from the plug-in location (see Fig. 22).
ȩ
Unscrew the cable connectors
2
with a TX 10 wrench.
ȩ
Separate the cable connector and remove any other cable connectors
present.
Note:
The screw-on cable connector
2
must be replaced by a new
cable connector every time the L2 lamp is replaced.
Note:
The lamp cable of the new L2 lamp must be guided underneath
the lamp holder (lamp housing) during the mounting.
Fig. 22
2
1
4
3
L2 / Ersatzlampe
L1 / Standardlampe
Disconnecting the lamp cable – lamp type B
Procedure