Lamp replacement and calibration - Corona process | Service manual |
EN_40_070_0010I
Replacing halogen lamps
Lamp replacement and calibration Corona process
2
Reinstalling the lamp holder
Carefully place the lamp holder (see Fig. 24 /
1
) on the base of the lamp – do
not touch the light guides in the process (see Fig. 24 /
2
).
Ensure that the black marking on the lamp socket is pointing upwards in its
installation position (see Fig. 27 /
4.1
).
Reattach the lamp holder to the base of the lamp (see Fig. 23 /
3
) using the
four screws (see Fig. 23 /
2
).
L1 lamp / Default lamp:
Guide the lamp cable for lamp L1 back through the device to the circuit
board as shown in Fig. 22.
Reattach the lamp cable to the cable bracket (see Fig. 22 /
1
) and lock it.
Reconnect the plug of the lamp cable to the board socket plug
(see Fig. 22 /
4
).
Replace all previously removed cable ties. Ensure that all cables are firmly
positioned.
L2 lamp / Spare lamp:
Guide the lamp cable for the L2 lamp to the circuit board as shown in Fig. 22.
Note:
The lamp cable of the L2 lamp may also be guided underneath the lamp
holder (lamp housing).
Reconnect the plug of the lamp cable to the board socket plug
(see Fig. 22 /
3
).
Replace the cable bracket (Fig. 22 /
2
) with a new one.
Replace all previously removed cable ties. Ensure that all cables are firmly
positioned.