Service manual | Lamp replacement and calibration - Corona process
Determining the device version
Lamp replacement and calibration Corona process
3 Determining the device version
1
Connecting the device to test computers and power supply
Connect the device to the measuring computer via the ethernet cable.
Establish the power supply for the device via the lab power supply.
2
Opening OSIS software
Start the
OSISManagementConsole.exe
under:
⇒
C:\Program Files (x86)\ZEISS\InProcess
⇒
C:\Program Files (x86)\ZEISS\InProcess\OSIS\ManagementConsole\
3
Selecting the device
Select the
Corona process
measuring system.
Fig. 3 Selecting the device
4
Enter the IP address of the measuring system or carry out a scan
Under
Hardware settings
>
IP address,
enter the IP address of your device or
carry out an address scan via the
Scan
button.
Ensure that you are working within the same IP address area when
connecting the device to the test computer.
Procedure
up to InProcess Version 2.9:
starting from InProcess
Version 2.10: