Service manual | Lamp replacement and calibration - Corona process
Replacing halogen lamps
Lamp replacement and calibration Corona process
8
Removing the base of the device from the device cover
Loosen the four hexagonal screws
1
of the housing with a 8 mm spanner.
(Phillips screws may be installed in older device versions. Loosen these using a
PH 2 Phillips screwdriver.)
Caution
RISK OF COMPONENT DAMAGE
Risk of damaging the FBG filter board while removing the device base
from the cover.
Carefully lift up the base and watch out for stuck parts.
Carefully lift the device base
2
out of the device cover
4
by the handle
3
. Avoid jamming any parts whilst removing.
Rotate the base of the device by 180° – the internal assemblies are now on
top and the handle on the bottom – and carefully place it at an ESD work
station.
Fig. 15
3
1
2
4
Removing the base of the device from the device cover