Service manual | Lamp replacement and calibration - Corona process
Sensor calibration
Lamp replacement and calibration Corona process
7
White calibration
Place the standard holder
2
with the white standard
3
at position
P1
.
Loosen the clamping screw
5
and slide the white standard along its sliding
axis
1
to the measurement window
4
of
Corona process
until touching it.
Fig. 37
4
3
2
1
5
6
Positioning the white standard
Keep the white standard
3
in this position by tightening the
clamping screw
5
.
Fig. 38 Continuing the white calibration
Press the
Continue
button to continue the process.
Sliding axis of the white /
black standard
Standard holder
White standard
Measurement window
Clamping screw
Adjustment screws
1
2
3
4
5
6