Service manual | Lamp replacement and calibration - Corona process
Sensor calibration
Lamp replacement and calibration Corona process
9
Recording spectra for distance correction
Switch back to the white standard and place the holder on position
P1
.
Ensure that the white standard touches the measurement window.
Keep the position by tightening the clamping screw (see Fig. 37).
Fig. 41 Continuing the distance correction
Press the
Continue
button to continue the process.
Position the standard holder including white standard at position
P2
.
Press the
Continue
button to continue the process.
Follow the instructions in the software and position the standard holder
with the white standard until position
P12
is reached.