Service manual | Lamp replacement and calibration - Corona process
Sensor calibration
Lamp replacement and calibration Corona process
Connect the ethernet cable and the power supply cable to the measuring system.
Start the
OSISManagementConsole.exe
under:
⇒
C:\Program Files (x86)\ZEISS\InProcess
⇒
C:\Program Files (x86)\ZEISS\InProcess\OSIS\ManagementConsole\
If not already done so, open the menu item
New
and select the
Corona process
measuring system.
Fig. 30 Selecting the measuring system
3
Select the lamp to be calibrated
Right-click
Corona process,
then select
Calibration of the default
lamp
or
Calibration of the spare lamp
.
Fig. 31 Selecting the lamp
It may take a few seconds to initialize the
Corona process
.
up to InProcess Version 2.9:
starting from InProcess
Version 2.10: