Lamp replacement and calibration - Corona process | Service manual |
EN_40_070_0010I
Sensor calibration
Lamp replacement and calibration Corona process
10
Saving the distance correction
When the calibration of the default lamp is completed the correction file will
be saved in the following folder: “C:\ProgramData\ZEISS\OSIS\Device\...”
Note:
Transmit this *.zip file to ZEISS Service, see step “14 Transmit *.ini files
Fig. 42 Saving the distance correction
Click
OK
to complete the process.
11
Calibrating the spare lamp
Now calibrate the spare lamp by right-clicking
Corona process
and
selecting
Calibration of the spare lamp
.
Fig. 43 Selecting the spare lamp
Repeat steps 4 to 10 for the spare lamp.