41
EN_40_070_0010I
|
Service manual | Lamp replacement and calibration - Corona process
Sensor calibration
Lamp replacement and calibration Corona process
12
Re-read the device data
Re-read the device data (as described in Chapter “3 Determining the device
version” on page 9).
13
Creating the new *.ini file
Under “Device info”, click on the
Export details
button. An *.ini file is created.
Save this file.
Fig. 44 Creating the new *.ini file
14
Transmit *.ini files and *.zip file
Transmit:
– the second *.ini file created in step 13,
– the first *.ini file created during preparation of the lamp replacement
(see also chapter 2, step “9 Generate *.ini file” on page 13) and
– the *.zip file created in step 10 (distance correction)
to Service at Carl Zeiss Spectroscopy GmbH.
⇒
15
Reinstalling the Corona process
Remove the
Corona process
measuring system from the calibration device.
In accordance with the specifications from the corresponding user manual,
reinstall the
Corona process
measuring system at the original location of use.
Concluding measures