Service manual | Lamp replacement and calibration - Corona process
Replacing halogen lamps
Lamp replacement and calibration Corona process
Ensure that the new halogen lamp is aligned so that the black marking on the
lamp socket
4.1
is pointing upwards in its installation position.
Fig. 27
4.1
6
5
Aligning the halogen lamp
Place the clamping ring
3
on the halogen lamp.
Reinsert the three screws in the lamp holder
1
using a TX 10 screwdriver
2
.
Fig. 28
1
2
3
Installing the clamping ring
Tighten the clamping ring with the three screws. Ensure that no gap remains
between the clamping ring and the lamp holder.