81
81
IM 54051-E
●
External probe (optional) specifications
Standard needle probe (Model: 900 10) *NOTE
Measuring range: -20 to 200
°
C (-4 to 392
°
F)
Range
Accuracy
-20
≤
C
≤
0 (-4
≤
F
≤
32)
±
0.5
C (TYP) (
±
0.9
F (TYP))
0<
C <100 (32<
F <212)
±
0.5
C (MAX) (
±
0.9
F (MAX))
100
≤
C <
150 (212
≤
F <302)
±
1.0
C (TYP) (
±
1.8
F (TYP))
150<
C
≤
200 (302<
F
≤
392)
±
2.0
C (TYP) (
±
3.6
F (TYP))
High speed needle probe (Model: 900 11) *NOTE
Measuring range: -20 to 200
C (-4 to 392
°
F)
Range
Accuracy
-20
≤
C
≤
150 (-4
≤
F
≤
302)
±
1.5
C (TYP) (
±
2.7
F (TYP))
150<
C
≤
200 (302<
F
≤
392)
±
2.5
C (TYP) (
±
4.5
F (TYP))
Surface probe (Model: 900 12) *NOTE
Measuring range: -20 to 200
C (-4 to 392
°
F)
Range
Accuracy
-20
≤
C
≤
150 (-4
≤
F
≤
302)
±
1.5
C (TYP) (
±
2.7
F (TYP))
150<
C
≤
200 (302<
F
≤
392)
±
2.5
C (TYP) (
±
4.5
F (TYP))
Rounded end probe (Model: 900 13) *NOTE
Measuring range: -20 to 200
C (-4 to 392
°
F)
Range
Accuracy
-20
≤
C
≤
0 (-4
≤
F
≤
32)
±
0.5
C (TYP) (
±
0.9
F (TYP))
0<
C <100 (32<
F <212)
±
0.5
C (MAX) (
±
0.9
F (MAX))
100
≤
C <
150 (212
≤
F <302)
±
1.0
C (TYP) (
±
1.8
F (TYP))
150<
C
≤
200 (302<
F
≤
392)
±
2.0
C (TYP) (
±
3.6
F (TYP))
Emission thermo probe (Model: 900 03)
Measuring range: -20 to 200
C (-4 to 392
°
F)
Accuracy
: Measurement temperature
±
1%
±
1
C or
±
3
C of the indication whichever is greater.
*NOTE : The external probe assumes the condition where it is dipped in the liquid being mixed to a half
dimension of the temperature measurement section.