42
42
4 Setting Up with the Personal Computer
IM 54051-E
7.
Click in the “Measuring period” column.
The cursor moves to a “Measuring period” field.
8.
Enter the measurement period with numeric characters, a backslash, and colons
(Example: 003/20:30:00).
[TIP]
Use numeric characters, a backslash, and colons to enter a measurement period. The input range is:
ddd < 1000, hh < 24, mm < 60, and ss < 60, where “ddd” denotes the number of days, “hh” denotes
hours, “mm” denotes minutes, and “ss” denotes seconds.
All the period data must be twelve characters or less (i.e. “ddd:hh:mm:ss”. d, h, m, and s are optional
numeric characters).
The specified measurement period has been set.
If multiple tags have to be set, select a “Name” cell from the next line, then repeat steps 2 through 8.