17
IM 54051-E
3 Setting Up with the Main Unit
17
17
Preparation for
Measurement
9.
Enter the lower-limit temperature.
Lo Alarm
+ 0 2 0 . 0
°
C
[TIP]
In order to set a negative value, press either
the [
▼
] or [
▲
] key. The “+” symbol on the
screen will change to a “-” symbol.
10
.Press the [SET] key.
The specified lower-limit temperature is set and the
“Turn” screen will appear on the screen.
Turn
OFF
11
.Use either the [
▼
] or [
▲
] key to select
whether the operation of the alarm func-
tion is to be inverted. If ON is selected,
the alarm function will operate under the
inverse condition.
12.
Press the [SET] key.
The range in which the alarm sounds will be displayed.
Then, the screen for confirming the settings will ap-
pear.
Change?
Yes1/No0
13
.Press the [1] key.
The range in which the alarm function is active will
be set for all tags as a batch.
Press the [ESC] key repeatedly to return to the Home
screen.
[TIP]
If another alarm function is batch-set for all
tags after one alarm function has been batch-
set, only the latest one will be effective.
-End of procedure-
■
Registering the measurement
operator name
Home screen
18:54:27
222
25.9
°
C
1.
Press the [FUNC], [8], and [3] keys, in
this order.
F83
E n t r y
2.
Press the [SET] key.
A screen for registering an operator name will ap-
pear.
1
S m i t h
3.
Press the [
▼
] key repeatedly until a
number assigned to the operator name
to be registered appears on the screen.
2
[TIP]
Each operator name is assigned a unique reg-
istration number. A maximum of ten operators
can be registered in the main unit.
See Also
For information about entering a character or
number, refer to “6 How to Input Characters”
in the “ Information Required before Use” part.