14
IM 54051-E
14
14
3 Setting Up with the Main Unit
7.
Press the [SET] key.
A screen for confirming the entered contents will ap-
pear.
Insert?
Yes1/No0
8.
Press the [1] key.
A new tag will be created.
Convey
0
Press the [ESC] key repeatedly to return to the Home
screen.
-End of procedure-
■
Setting the alarm function for
each tag
The alarm function can be set according to
different tag conditions, or the same alarm
function can be set for all tags.
The following procedure is used to set the
alarm function for different conditions.
See Also
To set the same alarm function for all tags, pro-
ceed to “
■
Batch-setting the alarm function”.
Home screen
18:54:27
222
222
25.9
°
C
1.
Press the [FUNC], [7], and [2] keys, in
this order.
The “1File” screen will appear on the LCD.
F72
1 F i l e
2
. Press the [SET] key.
A screen for selecting a tag name will appear.
Convey
0
3
. Press the [
▼
] key repeatedly until a tag
name for which the alarm is to be set
appears on the screen.
4
. Press the [SET] key.
A screen for setting the alarm HIGH function ON/
OFF will appear on the LCD.
Hi Alarm
ON
5
. Use either the [
▼
] or [
▲
] key to set the
upper-limit temperature alarm function
ON/OFF.
6
. If ON has been selected, press the [SET]
key.
The “Hi Alarm” setting screen will appear on the
screen.
If OFF has been selected in step 5, this screen will
not appear. In this case proceed to step 8.
Hi Alarm
+ 0 0 0 . 0
°
C