3 Setting Up with the Main Unit
11
11
IM 54051-E
Preparation for
Measurement
3
Setting Up with the Main Unit
3.1
Setups required to use the collector function
(Setting the tag name, input channel, alarm, and operator name)
Before using the collector function, the following tag setups must be made.
■
Setup contents
●
Tag name
Put a title for each tag.
●
Input channel
Thermo-Collector has multiple input channels and probes for measuring temperatures of
substances in a vapor, solid, or liquid phase. Since the probe to be used depends on the
measuring object, it is necessary to set the available input channel on a tag in advance.
The following table shows the probe types and their corresponding input channels.
Probe type
Input channel
Standard needle probe
A
High speed needle probe
Surface probe
Rounded end probe
Built-in sensor
C
Emission thermo probe
D
●
Alarm
Set a limit temperature at which the alarm sounds.
•
ON/OFF of the upper-limit alarm and upper-limit temperature
Make the following settings to sound an alarm when the temperature rises above a preset limit.
The upper-limit temperature cannot be set if the alarm function for the upper limit is set to OFF.
Temperature
Setting the upper-limit alarm to ON and the upper-limit temperature to 70
°
C
70
°
C
: Range in which the alarm sounds.