4-4
IM 701510-01E
4.3
Initializing Settings
1.
Press the
INITIALIZE key
to display the initialization execution
menu.
2.
Press the “
INITIAL EXEC
” soft key to execute initialization.
Explanation
Default settings related to the vertical and horizontal axes
CH1 to CH4
All channels ON (CH1/CH2 for DL1520/DL1520L)
V/DIV
50 V/div
POSITION
The horizontal position is set to 0div. For the vertical position, refer to
the figures (Two channels, Four channels) given under title
“POSITION” in section 4.2.
INPUT
Input coupling: DC, Probe attenuation: 10:1, Inversion: OFF
TIME/DIV
1 ms/div
TIMEBASE
INT
Default settings related to trigger
MODE
AUTO
FUNCTION
EDGE
LEVEL/SOURCE
Trigger source: CH1, Trigger slope: “
”,Trigger level: 0 V
COUPLING
Trigger coupling: AC, HF rejection: OFF
POSITION
Center of the waveform display frame
DELAY
Delay time: 0 s
HOLD OFF
Mode: OFF, Hold-off time: MIN (0.2
µ
s)
Default settings related to other settings
ACQUISITION
Mode: NORMAL
LENGTH (for DL1520L/DL1540L)
Record length: 10K
FILTER
Smoothing: OFF, Band limit: FULL
DISPLAY
Interpolation: SINE, X-Y display: OFF, Accumulate: OFF
DISPLAY FORM
Graticule: GRID, Scale: ON, % Marker: OFF, Waveform information:
ON, Intensity: TEXT (9)
SHIFT
OFF (LED is not lit)
Settings which cannot be initialized
In the following cases, the settings which were in effect before initialization will be retained
even if initialization is performed.
• Date and time
• Displayed waveforms/setting parameters
• GO/NO-GO zone waveforms
stored in the internal memory
• Communications addresses
Points to note when initializing
• Snapshot waveforms will be deleted when initialization is performed.
• Operations such as measurements, computations, etc. will be halted.
• Initialization cannot be performed during GO/NO-GO.
Keys and Procedure
Soft keys
INITIALIZE key