10-6
IM 701510-01E
10.3 Real-Time Printing using the Optional Built-in
Printer
1.
Press the
SHIFT + COPY (MENU) key
to display the hardcopy
setting menu. “
CENTR
” menu is also provided with DL1520/
DL1520L with suffix code -C3.
2.
Press the “
PRNTR
” soft key to display the printout type selection
menu.
3.
Press the “
REAL
” soft key.
Outputting to the printer
4.
Press the
COPY key
to execute the printing.
Aborting the printing
5.
After having performed step 4, press the
COPY key
again to abort the
printing.
Explanation
Chart speed
The chart speed depends on the time axis setting as shown in the table below.
Time axis setting
Chart Speed
500 ms/div
16.67 mm/s
1 s/div
8.33 mm/s
2 s/div
4.17 mm/s
5 s/div
1.67 mm/s
10 s/div
0.83 mm/s
20 s/div
0.42 mm/s
50 s/div
0.17 mm/s
Points to note when printing real-time
Before carrying out real-time printing, verify the following settings.
• The trigger mode should be set to Auto mode or Auto Level mode;
• The history memory function should be set to OFF;
• The time base should be selected INT;
• For DL1520L/DL1540L, the record length must be set to 10 K or 100 K.
• The time axis setting should lie within 500 ms/div to 50 s/div;
• Waveform acquisition should be started.
During real-time printing, only the execution/abortion operations are valid.
During real-time printing, the cursor measurements and waveform parameter measurements
will be halted.
When you press the START/STOP key during real time printing, data acquisition will stop
and therefore real-time printing will also stop. Pressing the START/STOP key again results
in resuming data acquisition and therefore real-time printing.
Real-time printing is not possible during ZOOM/MainZoom display, X-Y display or when
loaded/recalled waveforms are being displayed.
Keys and Procedure
COPY(MENU) key
SHIFT key
Soft keys