Relay functions
5.3.1
General information
The analyzer has 12 relays that are independent of each other and can
be linked to the analyzer module. The following functions are possible
for a relay output:
Proportional controller (Frequency controller or Pulse-width control-
ler, see section 5.3.2)
Limit indicator (see section 5.3.3)
Indicator contact for a monitored event (see section 5.3.8)
Relay action
The following actions of the relay can be laid down in the
Action
setting:
Note
In the case of monitoring functions, select
Normally closed
for the relay
action.
5.3.2
Proportional controller: Basic information
In the case of proportional control, a relay switches cyclically on and off
in a defined measured value range (proportional range). At the same
time, the relay switches with a:
duration of operation that corresponds to the measured value
(pulse-width controller, see page 5-30) or
switching frequency (frequency controller, see page 5-32).
Proportional controllers can be used in the following way:
Regulation with a single relay:
a regulating range with
Start value
and
End value
is defined. No
regulation takes place above and below the range of regulation (see
page 5-30).
Regulation with two relays:
a regulating range with
Start value
and
End value
is defined for each
relay. One relay regulates in the upper range of regulation and a
further relay in the lower range of regulation (see page 5-30).
Settings
Explanations
Normally closed
The relay should open if any event occurs.
Normally open
The relay should close if any event occurs.
Содержание TresCon UNO A111
Страница 2: ...2 ba43134e08 05 2017 TresCon UNO Copyright 2017 Xylem Analytics Germany GmbH Printed in Germany...
Страница 6: ...Inhaltsverzeichnis TresCon UNO 0 4 ba43134e08 05 2017 9 1 Abbreviations 9 1 9 2 Index 9 1...
Страница 18: ...Safety TresCon UNO 2 4 ba43134e08 05 2017...
Страница 56: ...Initial commissioning TresCon UNO 4 4 ba43134e08 05 2017...
Страница 133: ...TresCon UNO Operation 5 77 ba43134e08 05 2017 9 Select Yes or No with 68 and confirm with 3...
Страница 167: ...TresCon UNO Maintenance 6 11 ba43134e08 05 2017 6 Leave the menu with 4...
Страница 202: ...Appendix TresCon UNO 8 24 ba43134e08 05 2017...
Страница 207: ......