Operation
TresCon UNO
5 - 98
ba43134e08
05/2017
Configuring the
permeate switch
The configuration of the permeate switch on the software side is made
in the
Permeate switch
menu.
Note
The designations, Sample 1 and Sample 2 refer to the installation of
the overflow vessels according to section 3.7.1 I
Configure the permeate switch as follows:
Parameter
Values
Valve control
1, 2, 3 (number of the valve con-
nection on the terminal strip)
Measuring duration of sample 1, 2
10 ... 999 min
1
Measuring mode
Call up the
General Parameters
menu: Simultaneously press
the keys,
4
and
2
.
2
Select the
Permeate switch
menu item with
68
and
confirm with
3
. The prompt
Presentation of data/Change
appears
.
3
Select the
Change
mode with
68
and confirm with
3
.
4
Input the PIN digit-by-digit with
68
. Change the digit
with
57
and confirm with
3
. The
Permeate switch
menu appears on the display.
5
Select the parameter to be changed with
68
and switch
into input mode with
3
.
6
Enter the required value with
68
and change the digit
with
57
.
Содержание TresCon UNO A111
Страница 2: ...2 ba43134e08 05 2017 TresCon UNO Copyright 2017 Xylem Analytics Germany GmbH Printed in Germany...
Страница 6: ...Inhaltsverzeichnis TresCon UNO 0 4 ba43134e08 05 2017 9 1 Abbreviations 9 1 9 2 Index 9 1...
Страница 18: ...Safety TresCon UNO 2 4 ba43134e08 05 2017...
Страница 56: ...Initial commissioning TresCon UNO 4 4 ba43134e08 05 2017...
Страница 133: ...TresCon UNO Operation 5 77 ba43134e08 05 2017 9 Select Yes or No with 68 and confirm with 3...
Страница 167: ...TresCon UNO Maintenance 6 11 ba43134e08 05 2017 6 Leave the menu with 4...
Страница 202: ...Appendix TresCon UNO 8 24 ba43134e08 05 2017...
Страница 207: ......