5.3.9
Display and recording of relays
Display of the
relays in the measured
value display
The allocated relays appear on the display as additional information to
each measured value.
Fig. 5-16 Display: Display of the relays
Below the measured value, all the linked relays K1...n are displayed. A
relay that is being active is displayed in reverse video.
Recording
If the
Messages
record mode is activated, a record is output (example
Fig. 5-17 Example of a Relay parameters record
Allocated
relays
Relay parameters 1
-----------------------------------------
PO4-P outlet activation
Function: Pulse-width controller
Cmin: 0.50 mg/l PO4-P
Cmax: 1.50 mg/l PO4-P
tmin: 5 s
tmax: 100 s
Action: Normally open
-----------------------------------------
TresCon>