Operation
TresCon UNO
5 - 56
ba43134e08
05/2017
Settings
* A limitation for system errors can be defined with the current limita-
tions
Imin
and
Imax
. This can, for example, prevent the dosage of inad-
missibly high quantities of precipitation agents.
Setting
Selection/Values
Explanation
Recorder
0 - 20 mA
4 - 20 mA
Signal range of the
recorder output
Cnominal
within the measu-
ring range
(depending on the
analyzer module)
Nominal value
Xp
-99.9 mA/mg/l
...
+99.9 mA/mg/l
Proportional range of the
controller.
Negative values result in
a negative characteristic
curve.
Tv
off
1 ... 9998
Hold-back time: Integral
part of the controller
(
Off
= not effective)
Tn
off
1 ... 9998
Reset time: Differential
part of the controller
(
Off
= not effective)
Imin
any within the sig-
nal range of the
recorder
Lower current limitation *
Imax
any within the sig-
nal range of the
recorder
Upper current
limitation*
Note:
Distance between
Imin
and
Imax
:
at least 5 mA
Response with stop
any in the range
0 ... 20 mA
In the Stop mode, the
current output provides
the defined current
value.
Also in the case of
Auto-
Cal
,
AutoClean
etc.
Hold
The current at the output
remains unchanged.
Содержание TresCon UNO A111
Страница 2: ...2 ba43134e08 05 2017 TresCon UNO Copyright 2017 Xylem Analytics Germany GmbH Printed in Germany...
Страница 6: ...Inhaltsverzeichnis TresCon UNO 0 4 ba43134e08 05 2017 9 1 Abbreviations 9 1 9 2 Index 9 1...
Страница 18: ...Safety TresCon UNO 2 4 ba43134e08 05 2017...
Страница 56: ...Initial commissioning TresCon UNO 4 4 ba43134e08 05 2017...
Страница 133: ...TresCon UNO Operation 5 77 ba43134e08 05 2017 9 Select Yes or No with 68 and confirm with 3...
Страница 167: ...TresCon UNO Maintenance 6 11 ba43134e08 05 2017 6 Leave the menu with 4...
Страница 202: ...Appendix TresCon UNO 8 24 ba43134e08 05 2017...
Страница 207: ......