TresCon UNO
Maintenance
6 - 15
ba43134e08
05/2017
6.2
Disassembling the module for repair purposes
Attention
Depending on the analyzer module, the consumables in the wet chemi-
cal system can be harmful to health (e.g. corrosive or irritating).
Observe the relevant safety instructions in the operating manual of the
analyzer module and keep to the protective measures given there.
Disassembling the
analyzer module
Fig. 6-4
Pulling the tubes into the enclosure
1
Empty the wet chemical system of the analyzer module com-
pletely with the aid of the
Empty system
service routine. The
course is described in detail in the operating manual of the ana-
lyzer module.
2
Open the enclosure door.
3
Switch off the TresCon UNO. Bring the On/Off switch to the O
position.
4
Pull the permeate tube (pos. 1 in Fig. 6-4) off the overflow ves-
sel and pull it into the TresCon UNO enclosure.
5
Pull all the consumables tubes (pos. 2 in Fig. 6-4) off the screw
connectors of the consumables containers and pull them into
the TresCon UNO enclosure.
1 2
O
I
P1
Содержание TresCon UNO A111
Страница 2: ...2 ba43134e08 05 2017 TresCon UNO Copyright 2017 Xylem Analytics Germany GmbH Printed in Germany...
Страница 6: ...Inhaltsverzeichnis TresCon UNO 0 4 ba43134e08 05 2017 9 1 Abbreviations 9 1 9 2 Index 9 1...
Страница 18: ...Safety TresCon UNO 2 4 ba43134e08 05 2017...
Страница 56: ...Initial commissioning TresCon UNO 4 4 ba43134e08 05 2017...
Страница 133: ...TresCon UNO Operation 5 77 ba43134e08 05 2017 9 Select Yes or No with 68 and confirm with 3...
Страница 167: ...TresCon UNO Maintenance 6 11 ba43134e08 05 2017 6 Leave the menu with 4...
Страница 202: ...Appendix TresCon UNO 8 24 ba43134e08 05 2017...
Страница 207: ......