![wtw TresCon UNO A111 Скачать руководство пользователя страница 200](http://html1.mh-extra.com/html/wtw/trescon-uno-a111/trescon-uno-a111_operating-manual_1000694200.webp)
Appendix
TresCon UNO
8 - 22
ba43134e08
05/2017
Note
Due to internal processes, the RS485 interface may accept further
commands though the last command was not yet worked off. TresCon
signals this by the message, "
Last command still in process
".
NO2-N module:
Response:
<Date> <Time> <StdA> <EStdA> <StdB> <EStdB> <a0> <a1> <phi>
<CR> <LF>
Example:
14.10 13:49 0.00 9.13 0.40 433.08 -0.009 0.944 1.000
PO4-P module:
Response:
<Date> <Time> <StdA> <EStdA> <StdB> <EStdB> <a0> <a1> <phi>
<CR> <LF>
Example:
29.11 07:46 0.00 21.09 1.60 173.54 -0.221 10.495 1.000
4.)
Start measurement
Command:
T1PBRUN
Response:
TresCon> <CR> <LF>
5.)
Stop measurement
Command:
T1PBSTOP
Response:
TresCon> <CR> <LF>
6.)
Collective command
Command:
%!?T1PBGMDATA*
Response:
<Module> <Date> <Time> <Conc> <Unit> <Citation form> <M valid>
<Malfunction> <Status> <Error> <CR> <LF>
Example:
T1 04.04.03 15:42 10.93 mg/l NH4-N 1 0 M NO
Interpretation:
<M valid> states whether the measured value is valid: 1 = measured value valid, 0 =
measured value invalid.
<Malfunction> states whether a malfunction has occurred: 0 = no malfunction, 1 = mal-
function.
The following abbreviations are used for the <Status>: ACL (AutoClean),
AC (AutoCal), M (Measure), S (Stop) and B (other, e.g. pause with discontinuous
measuring operation). The statuses of this command do not always correspond to the
messages of PBGSTATUS.
With <Error>, errors and information are listed: No = no error or information, Ex = error
with the error code x (x = 1, 2, 3...), Hy = information with the information code y (y = 1,
2, 3...). The error and information codes serve as information for the WTW service per-
sonnel.
*For the NOx-N/SAK module, the nitrate value is returned using the %!?TxPBGMDATA
command. If the SAK or COD value shall be returned, the command can be extended
by "0" or "1":
%!?TxPBGMDATA1: returns <Conz> <Unit> <Citation form> of the SAK value
%!?TxPBGMDATA2: returns <Conz> <Unit> <Citation form> of the COD value
Содержание TresCon UNO A111
Страница 2: ...2 ba43134e08 05 2017 TresCon UNO Copyright 2017 Xylem Analytics Germany GmbH Printed in Germany...
Страница 6: ...Inhaltsverzeichnis TresCon UNO 0 4 ba43134e08 05 2017 9 1 Abbreviations 9 1 9 2 Index 9 1...
Страница 18: ...Safety TresCon UNO 2 4 ba43134e08 05 2017...
Страница 56: ...Initial commissioning TresCon UNO 4 4 ba43134e08 05 2017...
Страница 133: ...TresCon UNO Operation 5 77 ba43134e08 05 2017 9 Select Yes or No with 68 and confirm with 3...
Страница 167: ...TresCon UNO Maintenance 6 11 ba43134e08 05 2017 6 Leave the menu with 4...
Страница 202: ...Appendix TresCon UNO 8 24 ba43134e08 05 2017...
Страница 207: ......