![wtw TresCon UNO A111 Скачать руководство пользователя страница 58](http://html1.mh-extra.com/html/wtw/trescon-uno-a111/trescon-uno-a111_operating-manual_1000694058.webp)
Operation
TresCon UNO
5 - 2
ba43134e08
05/2017
5.1.1
Operating elements
The TresCon UNO is operated via the controller:
Controller
Fig. 5-1
Operating elements of the controller
Key functions
TresCon
Display
Measure key
Calibrate key
Return key
Coursor keys
Escape key
Key
Function per key stroke
Immediately starts an automatic calibration (AutoCal)
for the selected analyzer module. The current action is
terminated/stopped.
Switches immediately into the measuring mode. A rou-
tine is started (AutoClean or AutoCal) if required. The
current action is terminated/stopped.
Switches to the next menu level. Confirms inputs and
switches to the select/input mode for parameterization.
Switches to the next higher menu level. Calls up the
query
Store
during parameterization.
and,
at the same
time,
Switches from the measuring mode to the main menu
of the controller software (
General Parameters
).
Immediately stops the current program run (in the
measuring mode, if <<Stop appears in the status line)
and moves the cursor one position to the left.
Содержание TresCon UNO A111
Страница 2: ...2 ba43134e08 05 2017 TresCon UNO Copyright 2017 Xylem Analytics Germany GmbH Printed in Germany...
Страница 6: ...Inhaltsverzeichnis TresCon UNO 0 4 ba43134e08 05 2017 9 1 Abbreviations 9 1 9 2 Index 9 1...
Страница 18: ...Safety TresCon UNO 2 4 ba43134e08 05 2017...
Страница 56: ...Initial commissioning TresCon UNO 4 4 ba43134e08 05 2017...
Страница 133: ...TresCon UNO Operation 5 77 ba43134e08 05 2017 9 Select Yes or No with 68 and confirm with 3...
Страница 167: ...TresCon UNO Maintenance 6 11 ba43134e08 05 2017 6 Leave the menu with 4...
Страница 202: ...Appendix TresCon UNO 8 24 ba43134e08 05 2017...
Страница 207: ......