TresCon UNO
Maintenance
6 - 3
ba43134e08
05/2017
6.1.1
Testing the recorder
You can select default values for the recorder outputs. The currents
selected are immediately applied to the outputs as soon as they are
selected.
Setting up output
currents
Parameter
Values
I1 (output 1)
0 to 20 mA
I2 (output 2)
0 to 20 mA
I3 (output 3)
0 to 20 mA
1
Measuring mode
Call up the
Service
menu: Simultaneously press the keys,
4
and
2
. The request
Enter PIN
appears
.
2
Input the PIN digit-by-digit with
68
. Change the digit
with
57
and confirm with
3
. The
Service Mode
menu
appears
.
3
Select the
Recorder
menu item with
68
and confirm
with
3
. The
Recorder service
mode appears
.
4
Select the parameters with
68
and switch into input
mode with
3
.
5
Enter the required test current with
68
. Change to the
next digit with
57
and confirm with
3
.
Содержание TresCon UNO A111
Страница 2: ...2 ba43134e08 05 2017 TresCon UNO Copyright 2017 Xylem Analytics Germany GmbH Printed in Germany...
Страница 6: ...Inhaltsverzeichnis TresCon UNO 0 4 ba43134e08 05 2017 9 1 Abbreviations 9 1 9 2 Index 9 1...
Страница 18: ...Safety TresCon UNO 2 4 ba43134e08 05 2017...
Страница 56: ...Initial commissioning TresCon UNO 4 4 ba43134e08 05 2017...
Страница 133: ...TresCon UNO Operation 5 77 ba43134e08 05 2017 9 Select Yes or No with 68 and confirm with 3...
Страница 167: ...TresCon UNO Maintenance 6 11 ba43134e08 05 2017 6 Leave the menu with 4...
Страница 202: ...Appendix TresCon UNO 8 24 ba43134e08 05 2017...
Страница 207: ......