Installation
TresCon UNO
3 - 30
ba43134e08
05/2017
3.7.1
Installing the permeate switch
The permeate switch is screwed to the mounting element of the
TresCon UNO instead of the series overflow vessel.
Pre-assembling the
component
Replace the triple overflow vessel (left) of the permeate switch by the
simple overflow vessel of TresCon UNO (Fig. 3-25).
Fig. 3-25 Pre-mounting the permeate switch
1
Pull all tube endings (pos. 1) off the overflow vessel.
2
Unscrew the two knurled-head screws (pos. 2) and remove the
lid of the overflow vessel.
3
Unscrew the overflow vessel from the mounting plate.
4
Screw the simple overflow vessel on the mounting plate and
close it with the lid.
5
Attach labels to the overflow vessels (pos. 3): left = "Test
sample 2", right = "Test sample 1".
2
3
1
Probe 1
Probe 2
Содержание TresCon UNO A111
Страница 2: ...2 ba43134e08 05 2017 TresCon UNO Copyright 2017 Xylem Analytics Germany GmbH Printed in Germany...
Страница 6: ...Inhaltsverzeichnis TresCon UNO 0 4 ba43134e08 05 2017 9 1 Abbreviations 9 1 9 2 Index 9 1...
Страница 18: ...Safety TresCon UNO 2 4 ba43134e08 05 2017...
Страница 56: ...Initial commissioning TresCon UNO 4 4 ba43134e08 05 2017...
Страница 133: ...TresCon UNO Operation 5 77 ba43134e08 05 2017 9 Select Yes or No with 68 and confirm with 3...
Страница 167: ...TresCon UNO Maintenance 6 11 ba43134e08 05 2017 6 Leave the menu with 4...
Страница 202: ...Appendix TresCon UNO 8 24 ba43134e08 05 2017...
Страница 207: ......