3 Repairs
Copyright © WOODLEY EQUIPMENT COMPANY LTD.
90
Disassembly
1. Use a #2 (Ph2) Phillips screwdriver to disassemble the M3×8 cross-recessed pan-head
combination screw in the lower section of the shield cover to remove the cover; to take out the
screw, move slightly upwards for 3mm or so and apply force in a direction perpendicular to the
right clapboard. See Figure 3-22.
Figure 3-22 Dismantling and replacing the WBC bath assembly (1)
1 - Shield cover of the counting
bath
2 - Cross-recessed pan-head
combination screw
2. Pull out the peripheral fluid tubes attached to the WBC counting bath assembly. Pay attention to
handling the residual fluid; try to prevent it from flowing onto the other components in the
machine, and wipe clean if necessary.
3. Use a #2 (Ph2) Phillips screwdriver to disassemble the M3×8 cross-recessed pan-head
combination screw for grounding on the middle clapboard of the shield case cover and pull out
the grounding wire. See Figure 3-23.
Figure 3-23 Dismantling and replacing the WBC bath assembly (2)
1 - WBC counting chamber
assembly
2 - M3×8 cross-recessed
pan-head combination screw