3 Repairs
Copyright © WOODLEY EQUIPMENT COMPANY LTD.
74
Preliminary Steps
1. Turn off the power switch on the left side of the device and pull out the power cord plug from the
back panel of the device.
2. The left side panel has been disassembled. See
3.2.1 Disassembling the Left Side Panel
3. The right-side door is opened. See
3.2.2 Open the Right Side Door
Disassembly
1. Disassembly the back panel: use a #2 (Ph2) Phillips screwdriver to disassemble the three M3×6
cross-recessed pan-head combination screw on the back panel, and take off the back panel, as
shown in Figure 3-4.
Figure 3-4 Disassembling the back panel
1 - Right side door
2 - M3×8 cross-recessed
countersunk-head screw (×3)
2. Disassembling top cover: Open the right side door and disassemble 4 M3×8 cross-recessed
countersunk-head screws on both sides of top cover by #2 (Ph2) Philips screwdriver, the top
cover can be removed, as shown in Figure 3-5.
Содержание InSight V5
Страница 1: ...DMR MB 007 1 5...
Страница 126: ...7 Maintenance Inventory Copyright WOODLEY EQUIPMENT COMPANY LTD 120...
Страница 127: ...7 Maintenance Inventory Copyright WOODLEY EQUIPMENT COMPANY LTD 121...