5 Comprehensive Device Tuning
Copyright © WOODLEY EQUIPMENT COMPANY LTD.
110
d. Input the ipconfig command into the
cmd.exe
window popped out.
The interface shows similar content as follows:
The IPv4 address in the red box is the IP address of LIS workstation.
The IP address
192.168.8.44
of the LIS workstation shown as above is used as an example, real IP
should be in the same network segment with LIS server.
4. Click
OK
to save the settings.
5. Check if the connection is successful.
The LIS icon in the upper right side on the analyser screen turns from grey
to black
,
which indicates the InSight V5 Haematology Analyser software is connected to LIS successfully.
If the icon stays grey, the connection has failed. Please check if the IP address and port of LIS is
correct and reconnect as the steps above; if the problem still exists, please contact the hospital
network administrator.