5 Comprehensive Device Tuning
Copyright © WOODLEY EQUIPMENT COMPANY LTD.
108
5.5 LIS Connection
If the analyser needs to be connected to laboratory information system (hereinafter referred to as
LIS), you can complete the connection by following the steps in this section.
5.5.1 Installing LIS Workstation
1. Install LIS workstation and set instrument type and model.
2. Enter LIS workstation network setup interface after installation and set monitoring IP address
and port number.
Refer to
Description of LIS Communication Protocol for Haematology Analysers
to complete
the support of the LIS workstation to the LIS communication protocol.
5.5.2 Host Communication Settings
1. Use a network cable to connect the analyser to LIS local area network.
2. Log on the InSight V5 Haematology Analyser software; if the analyser is turned on, skip this step.
The whole process lasts for 4 to 12 minutes.
3. In the
Setup
interface, click
Host Communication
in the
Communication
selection to access
the Laboratory Information System (LIS) communication setting interface. See Figure 5-5.
Figure 5-5 Host Communication Settings