5 Comprehensive Device Tuning
Copyright © WOODLEY EQUIPMENT COMPANY LTD.
105
Figure 5-1 Position Adjustment
5.2 HGB Voltage Gain Setting
NOTE: The HGB voltage gain setting needs to be performed for the replacement of WBC bath, HGB
wire and/or main control panel.
1. In the
Setup
>
Meterage
screen, click
Gain Settings
.
2. Click the up/down adjustment button after the filling blank for HGB value and HGB background
voltage value will change accordingly within 4.5+/-0.3V.
3. Click
OK
(see Figure 5-2).
Figure 5-2 HGB voltage gain setting