3 Repairs
Copyright © WOODLEY EQUIPMENT COMPANY LTD.
81
Figure 3-13 Replacement of the sampling assembly in X- or Y-direction (2)
1 - Sampling assembly
2 - M4×10 cross-recessed pan-head
combination screw (×4)
3.5 Power Assembly Replacement
Purpose
The power assembly can be disassembled and replaced by following the procedures specified in this
section; however, in each case,
3.2 Preparatory Work before Repairs
begin.
Tools/Spare Parts
⚫
Phillips screwdriver
⚫
Power assembly with the same specifications
Disassembly
1. Use Phillips screwdriver to take out four M3×8 cross recessed pan-head combination screws on
the back of the device power supply assembly, then remove one M3×8 cross recessed pan-head
combination screw on the front, as shown in Figure 3-14.