6 Alarms and Solutions
Copyright © WOODLEY EQUIPMENT COMPANY LTD.
112
No.
Problem
Name
Problem Description
Solution
The motor fails to
move to the open
sampling position.
process is OK.
The horizontal motor
is busy.
Horizontal motor
timeout
The optocoupler of
the horizontal motor is
not working properly.
3
Vertical
motor
problem
The vertical motor
fails to leave its initial
position.
1. The user enters
Status
>
Sensor
screen and covers
the center of the vertical optocoupler with a piece of
paper. Check if the corresponding optocoupler status
shown in the screen is blocked. If it is, then the
optocoupler is working properly; if not, then refer to
3.4.2 Replacing the Optocoupler
the optocoupler.
3.4.3 Replacing the Sampling Assembly in
to disassemble the motor from the
sampling assembly and make sure motor wiring is
secure. If the wiring is OK, then disassemble the motor
and replace it with a new one. After the installation,
make the proper adjustments to the belt tension, then
go to the syringe self-test under
Service
>
Self-test
>
Syringe
. If the syringe is working, then it has been
successfully fixed.
The vertical motor
fails to return to its
initial position.
The motor fails to
move into position to
isolate the air
bubbles.
The motor fails to
move to the DIFF
bath position.
The motor fails to
move to the counting
bath position.
The motor fails to
move to the open
sampling position.
The vertical motor is
busy.
Vertical motor timeout
4
The CC
source
voltage is
abnormal
The CC source
voltage is abnormal.
3.9 Main Control Panel Replacement
replace the main control panel.