6 Alarms and Solutions
Copyright © WOODLEY EQUIPMENT COMPANY LTD.
116
No.
Problem
Name
Problem Description
Solution
No V5 LY1.
1. Check if V5 LY1 has air bubbles in inlet tubing or is
running out.
2. If there is no V5 LY1, replace with new V5 LY1. Click
the
Remove Error
button to automatically resolve
the error.
3. If the remaining volume of V5 LY1 is a lot, or the
error persists after replacing with new V5 LY1,
perform the next step.
4. Check whether the reagent detection module is
cracked. If it is cracked, replace the reagent
detection module component with a new one.
No V5 LY2.
1. Check if V5 LY2 has air bubbles in inlet tubing or is
running out.
2. If there is no V5 LY2, replace with new V5 LY2. Click
the
Remove Error
button to automatically resolve
the error.
3. If the remaining volume of V5 LY2 is a lot, or the
error persists after replacing with new V5 LY2,
perform the next step.
4. Check whether the reagent detection module is
cracked. If it is cracked, replace the reagent
detection module component with a new one.
16
Door or cover
open error
The right-side door is
open.
Check if the right-side door is open. The door should not
press against the microswitch. If the door is properly
closed, please replace the microswitch on the side door.
Optical assembly
cover is open.
Open the optical shield cover, then close it again to
check if the problem has been resolved; if not, replace
the optical microswitch.