2 Instrument Configuration
Copyright © WOODLEY EQUIPMENT COMPANY LTD.
43
Figure 2-36 Tubing connection of pinch valve
2. If the problem persists after the inspection, please check if the optical system is working properly.
For the common problems of the optical system and their solutions, see
2.3 Hardware System
The hardware system of the InSight V5 Haematology Analyser consists of the main control panel,
reagent testing panel, laser driver panel, optical preamplifier panel, front panel, reagent sealing
panel, etc. and its block diagram is shown in Figure 2-37.
Figure 2-37 Block Diagram of Hardware System
Main Control Panel
Reagent
closing Panel
Display screen
driver panel
Front panel (LED,
buzzer)
Other external connecting unit
(valve, pump, motor, sensor)
Laser driver
panel
Reagent
testing panel
Optical preamp
panel
Hardware maintenance principles apply to all the panels. All the panels have passed hardware tests
before leaving the factory, and any hardware problem found later may be mainly caused by power
problems. Thus, hardware maintenance primarily involves the power supply. Hardware problems
and maintenance methods for each panel are described in the subsequent sections.